boric acid

EU REACH-Ann17EU SVHCMolGod Score: Primary

5,99 €

Chemical reagent Kwas borowy (CAS 10043-35-3). Full encyclopedic card — classification, properties and safety data — below.

🔒 Demo mode — this item is not for sale.

Legal substance. dhscientific.com is a demonstration of the MOL-GOD platform — we sell nothing and no orders are fulfilled. Banned substances are blocked here automatically from the regulatory canon (compare e.g. Heptachlor).

Audit of your SDS — free → · Contact

MolGod_SDSCARD_1
REACH 2020/878
v2 · 22.07.2026
🧬 3D Molecule Visualizer
Loading molecule...
3D model Boric Acid, CAS 10043-35-3, molecular formula BH3O3, molar mass 61.84 g/mol
Chemical Overview: Boric AcidMolGod_OVERVIEW_1
Molecular formulaBH3O3
Molecular weight61.84 g/mol
Melting point171 °C
Density1.5 g/cm³
IUPAC nameboric acid
SMILESB(O)(O)O
InChIKeyKGBXLFKZBHKPEV-UHFFFAOYSA-N

Synonyms: BORIC ACID · Orthoboric acid · 10043-35-3 · Boracic acid · Borofax

Data sources: PubChem (NLM/NIH)
Last updated: 2026-11-10

SCIENTIFIC RESEARCH

[1]Europe PMC2026
et al.. (2026). "Effects of Dietary Salt and Boric Acid on Milk Quality in Savak Akkaraman Sheep.". https://doi.org/10.3390/ani16020233
[2]Europe PMC2026
(2026). "Impact of Boric Acid on the Stability of Hydrogen Peroxide in Aqueous Solution.". https://doi.org/10.1021/acsomega.6c03206
[3]Europe PMC2026
(2026). "Identification of a Riboflavin-Boric Acid Complex by Electrospray Ionization Mass Spectrometry.". https://doi.org/10.1007/s12011-026-05110-9
[4]Europe PMC2026
et al.. (2026). "Acute Boric Acid Poisoning With Chronic Renal Failure Successfully Treated With Continuous Hemofiltration.". https://doi.org/10.7759/cureus.103795
[5]Europe PMC2026
et al.. (2026). "Enhancing tuber quality, storage performance and yield response of potato to combined foliar application of zinc sulphate and boric acid.". https://doi.org/10.1038/s41598-026-46146-0
[6]Doaj2026
Xiaomei Zhang, Xinyu Chen, Yeersheng Jiangbaolati et al.. (2026). "Enhancing electrical conductivity of coal-derived graphite by boric acid catalysis: Correlating 2H/3R phase ratio with crystallite st
[7]Europe PMC2025
et al.. (2025). "Proteomic profiling of Escherichia coli cytoplasmic proteins under sublethal boric acid stress.". https://doi.org/10.55730/1300-0152.2745
[8]Europe PMC2025
et al.. (2025). "Unintentional boric acid exposure: a case report and boron level monitoring.". https://doi.org/10.1186/s12887-025-06254-8
📚 Scientific references (Chicago Author-Date) 20 refs · 4 baz

MOLEKUŁA Per-CAS bibliography (live from 13+ databases)

Sources: db:Europe PMC (11) · db:doaj (1) · db:openalex (8)

  1. db:Europe PMC et al.. (2026). "Effects of Dietary Salt and Boric Acid on Milk Quality in Savak Akkaraman Sheep.". https://doi.org/10.3390/ani16020233 →
  2. db:Europe PMC (2026). "Effects of boric acid on 'Muscat Hamburg' and 'Balıkçı Siyahı' grapevines under salt stress". https://doi.org/10.21203/rs.3.rs-9283096/v1 →
  3. db:Europe PMC (2026). "Impact of Boric Acid on the Stability of Hydrogen Peroxide in Aqueous Solution.". https://doi.org/10.1021/acsomega.6c03206 →
  4. db:Europe PMC (2026). "Identification of a Riboflavin-Boric Acid Complex by Electrospray Ionization Mass Spectrometry.". https://doi.org/10.1007/s12011-026-05110-9 →
  5. db:Europe PMC et al.. (2026). "Acute Boric Acid Poisoning With Chronic Renal Failure Successfully Treated With Continuous Hemofiltration.". https://doi.org/10.7759/cureus.103795 →
  6. db:Europe PMC et al.. (2026). "Enhancing tuber quality, storage performance and yield response of potato to combined foliar application of zinc sulphate and boric acid.". https://doi.org/10.1038/s41598-026-46146-0 →
  7. db:doaj Xiaomei Zhang, Xinyu Chen, Yeersheng Jiangbaolati et al.. (2026). "Enhancing electrical conductivity of coal-derived graphite by boric acid catalysis: Correlating 2H/3R phase ratio with crystallite structure.". PLoS ONE. https://doi.org/10.1371/journal.pone.0347483 →
  8. db:Europe PMC et al.. (2025). "Proteomic profiling of Escherichia coli cytoplasmic proteins under sublethal boric acid stress.". https://doi.org/10.55730/1300-0152.2745 →
  9. db:Europe PMC et al.. (2025). "Unintentional boric acid exposure: a case report and boron level monitoring.". https://doi.org/10.1186/s12887-025-06254-8 →
  10. db:Europe PMC et al.. (2025). "Boric acid-functionalized bentonite nanostructures attenuated paraquat-induced pulmonary injuries in rats.". https://doi.org/10.1038/s41598-025-27334-w →
  11. db:Europe PMC et al.. (2025). "Nuclear quantum dynamics of boric acid as probed by a thermal-to-epithermal neutron station.". https://doi.org/10.1038/s41598-025-29342-2 →
  12. db:Europe PMC et al.. (2025). "Kinetic and Thermodynamic Investigation into the Formation of Diol-Boric Acid Complexes Using NMR and Computational Analysis.". https://doi.org/10.1021/acsomega.5c07314 →
  13. db:openalex Giulia Montagner, Silvia Dominici, Simona Piaggi et al.. (2023). "Redox Mechanisms Underlying the Cytostatic Effects of Boric Acid on Cancer Cells—An Issue Still Open". Antioxidants. https://doi.org/10.3390/antiox12061302 →
  14. db:openalex Qiangqiang Liu, Yubo Chai, Ni Lin et al.. (2020). "Flame Retardant Properties and Thermal Decomposition Kinetics of Wood Treated with Boric Acid Modified Silica Sol". Materials. https://doi.org/10.3390/ma13204478 →
  15. db:openalex AtLee T. D. Watson, Vicki Sutherland, Helen Cunny et al.. (2020). "Postnatal Effects of Gestational and Lactational Gavage Exposure to Boric Acid in the Developing Sprague Dawley Rat". Toxicological Sciences. https://doi.org/10.1093/toxsci/kfaa061 →
  16. db:openalex Hasan Türkez, Fatime Geyikoğlu, Suat Çolak. (2011). "The protective effect of boric acid on aluminum-induced hepatotoxicity and genotoxicity in rats". TURKISH JOURNAL OF BIOLOGY. https://doi.org/10.3906/biy-0902-11 →
  17. db:openalex Erhan Ayşan, Fikrettin Şahi̇n, Dilek Telci et al.. (2011). "Body Weight Reducing Effect of Oral Boric Acid Intake". International Journal of Medical Sciences. https://doi.org/10.7150/ijms.8.653 →
  18. db:openalex Hasan Türkez, Fatime Geyikoğlu. (2010). "Boric acid: a potential chemoprotective agent against aflatoxin b1 toxicity in human blood". Cytotechnology. https://doi.org/10.1007/s10616-010-9272-2 →
  19. db:openalex Amin Rostami, Jamal Akradi, Firoz Ahmad‐Jangi. (2010). "Boric acid as cost-effective and recyclable catalyst for trimethylsilyl protection and deprotection of alcohols and phenols". Journal of the Brazilian Chemical Society. https://doi.org/10.1590/s0103-50532010000800026 →
  20. db:openalex (1987). "NTP Toxicology and Carcinogenesis Studies of Boric Acid (CAS No. 10043-35-3) in B6C3F1 Mice (Feed Studies).". PubMed.
📊 Physicochemical properties

Quick Reference

Formula: BH3O3
MW: 61.84 g/mol
CAS: 10043-35-3
Appearance: Colorless, transparent crystals or white granules or powder
Odour: Odorless
Odour: Odorless

Detailed Properties

Property Value Unit Conditions Source
Density (ρ) 1.5000 g/cm³ 20 °C (68 °F) PubChem PUG-View (2026)
Melting Point (mp) 171.00 °C (339.8 °F) 1 atm PubChem PUG-View (2026)
Boiling Point (bp) 572 °F at 760 mmHg (decomposes) (NTP, 1992) PubChem (NIH/NLM) ↗
Vapor Pressure 1.6X10-6 mm Hg at 25 °C (2.136X10-4 Pa); log P (in Pa) = 26.83 - 9094/T where T is deg K PubChem (NIH/NLM) ↗
Water Solubility 10 to 50 mg/mL at 66 °F (NTP, 1992) PubChem (NIH/NLM) ↗
Refractive Index (nD) Index of refraction: 1.337, 1.461, 1.462 PubChem (NIH/NLM) ↗
🔬 Advanced Properties

Chemical Identifiers

SMILES: B(O)(O)O
InChI: InChI=1S/BH3O3/c2-1(3)4/h2-4H
InChIKey: KGBXLFKZBHKPEV-UHFFFAOYSA-N

Data sources: PubChem, NIST Chemistry WebBook, CRC Handbook of Chemistry and Physics (103rd ed.)

Last updated: 2026-06-30

Regulatory status of the substance
This compound: is subject to restrictions for specific uses (REACH Annex XVII); appears on the SVHC Candidate List (REACH Art. 59). Inventories: EU/REACH-Ann17, EU/SVHC. Regulatory information — does not restrict purchase in this store.
🧮 Stoichiometry CalculatorMolGod_STOICH_1
🔍 External identifiersMolGod_EXTID_1
14 of 16 ID systems88%
DatabaseIdentifierActions
CAS Registry Number10043-35-3Open →
PubChem CID7628Open →
InChIKeyKGBXLFKZBHKPEV-UHFFFAOYSA-NOpen →
InChIInChI=1S/BH3O3/c2-1(3)4/h2-4H
SMILESB(O)(O)O
EC Number233-139-2Open →
DrugBankDB11326Open →
KEGG CompoundD01089Open →
HMDBHMDB0035731Open →
ChemSpider7346Open →
MeSH UID (NLM)C032688Open →
UNII (FDA)R57ZHV85D4Open →
NSC Number (NCI)81726Open →
WikiData QIDQ187045Open →

Sources: PubChem (NIH), Wikidata SPARQL, KEGG, ChEMBL (EBI), CompTox CTX (EPA).

📡 Spectroscopy — CAS 10043-35-3MolGod_SPECHUB_MAIN
📊 Spectroscopic spectra databases — inline data 9 sources MolGod_SPECDB_2

Spectra are fetched on demand from 9 sources. Each spectrum is stored in our database — the next time it is opened there are zero requests to the external API. Download JCAMP-DX / CSV / PNG for every spectrum without searching.

IR IR (Infrared) — NIST WebBook
Public domain (US Federal)
▶ Click to load spectrum
🔗 Source
— points
📚 NIST Chemistry WebBook, SRD 69
MS (NIST) Mass Spectrum (EI) — NIST WebBook
Public domain (US Federal)
▶ Click to load spectrum
🔗 Source
— points
📚 NIST Standard Reference Database 1A
UV-Vis UV/Visible Absorption — NIST WebBook
Public domain (US Federal)
▶ Click to load spectrum
🔗 Source
— points
📚 NIST Chemistry WebBook, SRD 69
¹H NMR NMR (¹H, ¹³C) — NMRShiftDB
CC-BY-SA 4.0
▶ Click to load spectrum
🔗 Source
— points
📚 Steinbeck C et al. (2003) J. Chem. Inf. Comput. Sci. 43(1):10–16 DOI: 10.1021/ci025588g
MS (MoNA) MoNA — MassBank of North America
CC-BY 4.0
▶ Click to load spectrum
🔗 Source
— points
📚 MassBank of North America (UC Davis) DOI: 10.1002/jms.1777
IR/NMR/MS (SDBS) SDBS — Spectral Database for Organic Compounds (Japan AIST)
Free for non-commercial

Reference source — no public API. Open in an external database:

🔗 IR/NMR/MS (SDBS) →
📚 SDBSWeb: https://sdbs.db.aist.go.jp (AIST, Japan)
JP Monograph Japanese Pharmacopoeia — Monographs
Reference only

Reference source — no public API. Open in an external database:

🔗 JP Monograph →
📚 Japanese Pharmacopoeia 18th Edition (2021)
WHO INN WHO — International Nonproprietary Names
WHO Model Lists (free)

Reference source — no public API. Open in an external database:

🔗 WHO INN →
📚 WHO INN Programme
DOAJ DOAJ — Directory of Open Access Journals
OA journal index (mixed)

Reference source — no public API. Open in an external database:

🔗 DOAJ →
📚 DOAJ — doaj.org
🔬 Interactive spectra (live — NIST / MoNA / NMRShiftDB / SDBS) (2)

Data retrieved live from multiple sources (priority chain). JCAMP-DX / CSV / PNG available for download under each spectrum.

IR — Fourier-transform infrared

Loading IR — Fourier-transform infrared…

MS — Mass spectrometry (EI 70eV)

Loading MS — Mass spectrometry (EI 70eV)…

📐 Physical & Chemical Properties (DB) 3 fields MolGod Score: Reliable
Property Value Unit Conditions Source
Melting point 171.00 [5][6][11] °C 1 atm PubChem PUG-View (2026)
Water solubility 50.000 [5] g/L 25°C PubChem PUG-View (2026)
Density (ρ) 1.5000 [5] g/cm³ 20°C PubChem PUG-View (2026)
📚 Scientific references (Chicago Author-Date) (12 sources)
  1. Haynes, William M., ed. 2024. CRC Handbook of Chemistry and Physics. 105th ed. Boca Raton: CRC Press/Taylor & Francis. ISBN 978-1-032-55554-4. ↗
  2. National Institute of Standards and Technology. 2024. "NIST Chemistry WebBook, SRD 69." Gaithersburg, MD: NIST. Accessed 2025-01-01. ↗
  3. Yaws, Carl L. 2014. The Yaws Handbook of Physical Properties for Hydrocarbons and Chemicals. 2nd ed. Oxford: Gulf Professional Publishing.
  4. PubChem. 2024. "PubChem Compound Database." National Library of Medicine, National Institutes of Health. Accessed 2025-01-01. ↗
  5. Marrero, J., and R. Gani. 2001. "Group-Contribution Based Estimation of Pure Component Properties." Fluid Phase Equilibria 183–184: 183–208. ↗
  6. Joback, K. G., and R. C. Reid. 1987. "Estimation of Pure-Component Properties from Group-Contributions." Chemical Engineering Communications 57 (1–6): 233–243. ↗
  7. Sangster, J. 1997. Octanol-Water Partition Coefficients: Fundamentals and Physical Chemistry. Chichester: Wiley. ISBN 978-0-471-97397-3.
  8. Mannhold, Raimund, and Han van de Waterbeemd. 2001. "Substructure and Whole Molecule Approaches for Calculating Log P." Journal of Computer-Aided Molecular Design 15 (4): 337–354. ↗
  9. Perrin, Ditlev D., Boyd Dempsey, and E. P. Serjeant. 1981. pKa Prediction for Organic Acids and Bases. London: Chapman and Hall. ISBN 0-412-21090-5.
  10. Constantinou, Leonidas, and Rafiqul Gani. 1994. "New Group Contribution Method for Estimating Properties of Pure Compounds." AIChE Journal 40 (10): 1697–1710. ↗
  11. Ertl, Peter, Bernhard Rohde, and Paul Selzer. 2000. "Fast Calculation of Molecular Polar Surface Area as a Sum of Fragment-Based Contributions and Its Application to the Prediction of Drug Transport Properties." Journal of Medicinal Chemistry 43 (20): 3714–3717. ↗
🔄 Concentration unit converter LIVE MolGod_UNITCONV_1

Enter the Boric Acid concentration in any unit — the rest will be calculated automatically.

MW: 61.84 g/mol · IUPAC Gold Book ↗

⚗️ Conversion formulas + citations (per formula)
ConversionFormulaAccuracySource
% (w/v) ↔ molarityc (mol/L) = (% × 10) / MW±0.5% rel. when density ≈ 1.0 g/mLIUPAC (2019)
millimolar ↔ molarc (mol/L) = mM × 10⁻³ExactCohen ER, Cvitaš T, Frey JG, Holmström B, Kuchitsu K, Marquardt R, Mills I, Pavese F, Quack M, Stohner J, Strauss HL, Takami M, Thor AJ (2007)
molarity (mol/L)c = n/V = (m/MW)/V±0.1% (depends on MW precision)IUPAC (2019)
parts per million (mg/L) ↔ molarityc (mol/L) = ppm / (1000 × MW); equivalently ppm = mg/L for dilute aqueous±1% (density-independent for dilute solutions)IUPAC (2019)
mg/mL ↔ molarityc (mol/L) = (mg/mL × 1000) / MW / 1000 = mg/mL / MW × 1±0.2%Cohen ER, Cvitaš T, Frey JG, Holmström B, Kuchitsu K, Marquardt R, Mills I, Pavese F, Quack M, Stohner J, Strauss HL, Takami M, Thor AJ (2007)
g/L ↔ molarityc (mol/L) = (g/L) / MW±0.1% (depends on MW precision)Cohen ER, Cvitaš T, Frey JG, Holmström B, Kuchitsu K, Marquardt R, Mills I, Pavese F, Quack M, Stohner J, Strauss HL, Takami M, Thor AJ (2007)
mmol/L ↔ molarityc (mol/L) = mmol/L × 10⁻³ExactCohen ER, Cvitaš T, Frey JG, Holmström B, Kuchitsu K, Marquardt R, Mills I, Pavese F, Quack M, Stohner J, Strauss HL, Takami M, Thor AJ (2007)
Celsius ↔ KelvinT(K) = t(°C) + 273.15±0.01 K (ITS-90 scale)BIPM (Bureau International des Poids et Mesures) (2019)
Celsius ↔ FahrenheitT(°F) = T(°C) × 9/5 + 32±0.1 °FThompson A, Taylor BN (2008)
density-corrected % ↔ molarityc (mol/L) = (%w/w × ρ × 10) / MW, ρ in g/mL±0.1% when ρ known to 3 decimalsCohen ER, Cvitaš T, Frey JG, Holmström B, Kuchitsu K, Marquardt R, Mills I, Pavese F, Quack M, Stohner J, Strauss HL, Takami M, Thor AJ (2007)
📚 Bibliography (8 authoritative sources)
  1. Thompson A, Taylor BN (2008). Guide for the Use of the International System of Units (SI). NIST Special Publication 811 · DOI: 10.6028/NIST.SP.811-2008
    → Primary SI standard for US scientific usage
  2. Cohen ER, Cvitaš T, Frey JG, Holmström B, Kuchitsu K, Marquardt R, Mills I, Pavese F, Quack M, Stohner J, Strauss HL, Takami M, Thor AJ (2007). Quantities, Units and Symbols in Physical Chemistry — The IUPAC Green Book. RSC Publishing, 3rd ed. · DOI: 10.1039/9781847557889 · ISBN: 978-0-85404-433-7
    → Canonical IUPAC guide for chemistry quantities/units
  3. BIPM (Bureau International des Poids et Mesures) (2019). The International System of Units (SI), 9th edition. BIPM · ↗
    → International SI definitions (incl. redefined kilogram 2019)
  4. ISO/IEC (2022). Quantities and units — Part 1: General. International Organization for Standardization — ISO 80000-1:2022 · ↗
    → General rules for physical quantities and units
  5. ISO/IEC (2019). Quantities and units — Part 9: Physical chemistry and molecular physics. International Organization for Standardization — ISO 80000-9:2019 · ↗
    → Concentration / molality / amount-of-substance conventions
  6. Tiesinga E, Mohr PJ, Newell DB, Taylor BN (2021). CODATA recommended values of the fundamental physical constants: 2018. Rev. Mod. Phys. 93(2):025010 · DOI: 10.1103/RevModPhys.93.025010
    → Avogadro, gas constant, molar volume (2019 SI revision)
  7. IUPAC (2019). Compendium of Chemical Terminology — the IUPAC Gold Book (online). IUPAC · DOI: 10.1351/goldbook
    → Definitions of mass fraction, molality, normality, ppm, activity
  8. Mills IM, Cvitaš T, Homann K, Kallay N, Kuchitsu K (1988). Quantities, Units and Symbols in Physical Chemistry. Blackwell Scientific Publications, 1st ed. · ISBN: 0-632-01773-5
    → Historical predecessor of IUPAC Green Book
🧪 Solution Preparation Wizard WIZARD MolGod_PREP_1
① Select concentration
② Target volume
③ Solvent

Calculations per: IUPAC Gold Book ↗, Merck ↗

🛡️ Safety — CAS 10043-35-3MolGod_SAFEHUB_MAIN
Data limitations notice. The safety information on this page is for reference only and does not replace a full safety data sheet (SDS). Before using the product, consult the manufacturer's current safety data sheet and the GHS/CLP guidance. The CLP classification applies to the pure bulk substance, not to commercial formulations.

GHS/CLP classification — Regulation (EC) No 1272/2008 + UN GHS Rev. 9 (2021).

⚠️ Danger
GHS08 — Health hazard
GHS08 Health hazard

🚨 Hazard statements (H)

  • H360FD — May damage fertility. May damage the unborn child

🛡 Precautionary statements (P)

  • P203 — Obtain, read and follow all safety instructions before use

✓ Harmonised classification pursuant to Annex VI of the CLP Regulation (EC) 1272/2008 (official, binding classification). Index number: 005-007-00-2.

Reference (Chicago): European Chemicals Agency. "boric acid, Index No. 005-007-00-2." In Table 3 of Annex VI to Regulation (EC) No 1272/2008 (CLP Regulation), 23rd Adaptation to Technical Progress. Helsinki: European Chemicals Agency, 2026. https://echa.europa.eu/information-on-chemicals/annex-vi-to-clp.

Translations: CLP Regulation (EC) 1272/2008, Annexes III and IV. Data: PubChem/NLM.

📚 Consolidated scientific references — Chicago Author-Date 10 sources

References collected from all Safety Hub tabs. CAS: 10043-35-3 · PubChem ↗

  1. Parlament Europejski i Rada UE. 2008. "Rozporządzenie (WE) nr 1272/2008 w sprawie klasyfikacji, oznakowania i pakowania substancji (CLP)." Dz.Urz. UE L 353. [↗] GHS, Regulations
  2. United Nations Economic Commission for Europe (UNECE). 2021. "Globally Harmonized System of Classification and Labelling of Chemicals (GHS), Ninth Revised Edition." United Nations, Geneva. [↗] GHS
  3. Goldfrank, Lewis R., Robert S. Hoffman, Mary Ann Howland, et al.. 2019. "Goldfrank's Toxicologic Emergencies, 11th ed.." McGraw-Hill Education, New York. ISBN 978-1-25-985961-8. Pierwsza pomoc, Toksykologia
  4. National Institute for Occupational Safety and Health (NIOSH). 2023. "NIOSH Pocket Guide to Chemical Hazards (DHHS Publ. 2005-149)." U.S. Department of Health and Human Services / CDC, Cincinnati, OH. [↗] Pierwsza pomoc, PPE, Toksykologia
  5. European Committee for Standardization (CEN). 2016. "EN 374-1:2016 — Protective gloves against dangerous chemicals and micro-organisms." CEN, Brussels. [↗] PPE
  6. UNECE. 2023. "European Agreement Concerning the International Carriage of Dangerous Goods by Road (ADR 2025)." United Nations, Geneva. [↗] Utylizacja, Regulacje
  7. National Fire Protection Association (NFPA). 2022. "NFPA 400 — Hazardous Materials Code." NFPA, Quincy, MA. [↗] Magazynowanie
  8. Urben, P.G. (ed.). 2017. "Bretherick's Handbook of Reactive Chemical Hazards, 8th ed.." Butterworth-Heinemann / Elsevier, Oxford. [↗] Magazynowanie
  9. Ministerstwo Klimatu i Środowiska RP. 2023. "Baza danych o produktach i opakowaniach oraz o gospodarce odpadami (BDO)." Ministerstwo Klimatu i Środowiska, Warszawa. [↗] Utylizacja
  10. International Agency for Research on Cancer (IARC / WHO). 2024. "IARC Monographs on the Identification of Carcinogenic Hazards to Humans — List of Classifications." WHO, Lyon. [↗] Toksykologia

Tabs with their own references (Emergency, PPE, Storage, Waste) contain additional bibliographic entries within their respective sections.

📈 Analytical statistics (t-test · RSD · Grubbs · Q-Dixon) ICH Q2

Paste a series of replicate measurements (CSV, or one number per line). The calculator computes the mean, standard deviation and 95% CI, and detects outliers (Grubbs + Dixon Q).

Separator: comma, space, tab, new line. Minimum 3 measurements.
📐 Statistical formulas
  • x̄ = Σxᵢ / n — arithmetic mean
  • s² = Σ(xᵢ - x̄)² / (n-1) — sample variance
  • s = √s² — standard deviation
  • RSD% = (s / x̄) × 100% — relative standard deviation
  • CI₉₅ = x̄ ± t(0.05, n-1) × s / √n — Student's t
  • G = |xᵢ - x̄| / s — Grubbs' test
  • Q = |xsuspect - xnearest| / |xmax - xmin| — Dixon Q-test

Source: ICH Q2(R2) Validation of Analytical Procedures · ICH PDF ↗

🧪 Buffer Recipe Calculator UNIQUE

Choose a buffer from the list of 20 popular systems → enter the target pH → get an exact recipe with the masses to weigh out.

Step 1: Choose a buffer system

📜 Recipe history (last 10)
🔬 HPLC/GC methods (1 method)
📄
A quality-by-design optimized LC method for navigating degradation kinetics and quantification of favipiravir in the presence of degradation products and manufacturing impurities
HPLCBMC Chemistry202592% ✓CC-BYResearch method (specificity)
Column: C18, 5.0 μm
Phase: mobile phase was adjusted using a bench-top pH-2005 m (SELECTA, Barcelona, Spain)
Detection: MS
Flow: 1.00 mL/min
Inj.: 0 μL
Gradient: elution containing ACN and had an average run time of…
Ibrahim A, Farouk M, Alamir S, Salman B, Belal T, El Deeb S, et al. A quality-by-design optimized LC method for navigating degradation kinetics and quantification of favipiravir in the presence of degradation products and manufacturing impurities. BMC Chemistry. 2025;19:244. doi:10.1186/s13065-025-01610-2
Emergent viruses require effective treatment regimens, and the severe acute respiratory syndrome coronavirus-2 variants are still evolving. Favipiravir has emerged as an effective antiviral medication, especially after its popularity against COVID-2019. Despite its growing popularity, favipiravir has not yet been included in any pharmacopeia, and new dosage forms are being developed to improve its bioavailability for various infections. Therefore, studying its stability and developing methods capable of detecting its impurities and degradation products are essential. An isocratic high-performance liquid chromatography with photodiode array detection method was also designed using the Quality-by-Design approach and validated to determine favipiravir in presence of its hydrolytic degradation products, besides two major manufacturing impurities (namely, 3,6-dichloro pyrazine-2-carbonitrile and 6-fluoro-3-hydroxypyrazine-2-carbonitrile). The separation was achieved using a mobile phase of 25.0 mM phosphate buffer (pH 3.04): acetonitrile (92: 8, v/v) at a 1.0 mL min−1 flow rate. A Hypersil C18-BDS column (5.0 μm, 250.0 × 4.6 mm) was employed at room temperature and detection at 323.0 nm. The method had a linear range (5.0–100.0 µg mL−1) with a limit of detection and quantification of 0.51 and 1.54 µg mL−1. The drug's stability after forced degradation and at different pH levels was explored. Key results proved that FAV is most stable at pH 5.0, with calculated activation energies for the acidic and alkaline degradation processes being 53492.276 and 61896.899 kJ/mole, respectively. Moreover, the analysis of expired favipiravir tablets revealed the presence of the alkaline degradate, thereby demonstrating the efficacy of the method. The proposed method was evaluated for its greenness and blueness to demonstrate its ecological safety and practicality, scoring 0.65 in the Analytical Greenness (AGREE) metric and 85.00 on the Blue Applicability Grade Index (BAGI). It is the fi...
FavipiravirDegradation kineticsStability-indicating HPLCQuality-by-design (QbD)Green chromatographyImpurity profiling
📈 Method Validation (ICH Q2)

No validation data. Contact the method author.

Parameters per: ICH Q2(R2) ↗

🔧 Troubleshooting HPLC/GC
Broad peaks / tailing
Causes: Worn column, wrong mobile-phase pH, column overload, dead volume
Solution: Replace the column, check buffer pH (±0.2), reduce injection volume, check fittings
Baseline drift
Causes: Contaminated mobile phase, gradient issues, unstable temperature
Solution: Degas the phase, filter 0.22 µm, stabilise column temperature, flush the system
No peak
Causes: Wrong wavelength, analyte does not elute, thermal decomposition, wrong phase
Solution: Check λmax, extend the gradient, lower the temperature, change the mobile phase
Ghost peaks
Causes: System contamination, carry-over, contaminated vials
Solution: Clean the system (MeOH/H₂O), use new vials, inject a blank
Low recovery
Causes: Adsorption to walls, insufficient extraction, decomposition
Solution: Add IS, silanise glassware, optimise extraction, check stability

Sources: Snyder, Kirkland & Dolan ↗, Waters ↗

📊 HPLC method validation (ICH Q2(R1)) PARTIAL

3 metrics need experimental data

Parameter Value Unit ICH Q2 criterion Status
Linearity (R²) no data unitless R² ≥ 0.999 (≥0.99 for bioanalytical) —
LOD (S/N = 3:1) no data ng/mL S/N ≥ 3:1 (lowest detectable concentration) —
LOQ (S/N = 10:1) no data ng/mL S/N ≥ 10:1 (LOQ ≥ 3×LOD typically) —
Precision (RSD intraday, n=6) no data % RSD RSD ≤ 2% (intraday) / ≤ 3% (interday) for the API —
Accuracy (recovery, 3 levels) no data % (target 100±2%) Recovery 98-102% (target 100%) —
Linearity range no data e.g. 0.1-100 ng/mL Min. 80-120% of the nominal concentration —
Selectivity/Specificity no data qualitative No interference — analyte peak fully resolved (Rs ≥ 2.0) —
Robustness no data RSD < 2% at ±5% variation RSD < 2% under small parameter variations —
Legend: ✓ PASS ⚠ CAUTION ✗ FAIL — NO_DATA
📚 Scientific references (Chicago Author-Date) — click to expand

Analytical method validation standards — 4 independent sources (ICH + USP + AOAC + Snyder).

  1. International Conference on Harmonisation (ICH). 2005. Validation of Analytical Procedures: Text and Methodology Q2(R1). ICH Expert Working Group. [link ↗] — Gold-standard ICH guideline — accepted by EMA, FDA, MHLW, NMPA
  2. United States Pharmacopeia (USP) Convention. 2024. USP General Chapter <621> Chromatography. USP-NF 2024 ed. USP. [link ↗]
  3. AOAC International. 2016. Appendix F: Guidelines for Standard Method Performance Requirements. AOAC INTERNATIONAL. [link ↗] — AOAC SMPR — alternative to ICH Q2 for food/dietary supplements
  4. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. Introduction to Modern Liquid Chromatography. 3rd ed. John Wiley & Sons. ISBN 978-0-470-16754-0. https://doi.org/10.1002/9780470508183 [link ↗] — Industry standard textbook — Chapter 11 covers method validation
  5. Snyder, L. R., J. J. Kirkland, and J. L. Glajch. 1997. Practical HPLC Method Development. 2nd ed. Wiley. ISBN 978-0-471-00703-6. — Classic method-development reference (DryLab heritage).
  6. Rozet, Eric, et al.. 2013. Analysis of recent pharmaceutical regulatory documents on analytical method validation. https://doi.org/10.1016/j.chroma.2007.03.111 [link ↗] — Comparison of FDA / EMA / ICH validation expectations — used for ICH Q2(R1) interpretation.
  7. Heyden, Yvan Vander, et al.. 2009. Robustness of pharmaceutical liquid chromatographic methods. https://doi.org/10.1016/j.jchromb.2008.10.052 [link ↗] — Plackett-Burman design for robustness — basis of ICH Q2 §3.7.
  8. Dong, Michael W.. 2019. HPLC and UHPLC for Practicing Scientists. 2nd ed. Wiley. ISBN 978-1-119-31378-3. https://doi.org/10.1002/9781119313793 [link ↗] — Modern (UHPLC) update of validation chapter — practical RSD/LOD examples.
  9. Meyer, Veronika R.. 2010. Practical High-Performance Liquid Chromatography. 5th ed. Wiley. ISBN 978-0-470-68218-0. — European pharmacopeial perspective — complements USP/AOAC.
  10. Kazakevich, Yuri V., and Rosario LoBrutto, eds.. 2007. HPLC for Pharmaceutical Scientists. Wiley-Interscience. ISBN 978-0-471-68162-4. https://doi.org/10.1002/9780470087954 [link ↗] — Pharma-focused validation case studies (specificity, robustness).
  11. European Medicines Agency (EMA). 2011. Guideline on bioanalytical method validation EMEA/CHMP/EWP/192217/2009. EMA Committee for Medicinal Products for Human Use. [link ↗] — EMA bioanalytical companion to ICH Q2(R1) for clinical samples.

· ⚠ SVHC/REACH regulatory warnings ↑

🔧 HPLC troubleshooting — decision tree 6 common problems

Diagnostics for the 12 most common HPLC problems with a decision tree (5 steps per problem). Source: Snyder/Kirkland/Dolan 3rd ed. Chapter 17 + LCGC LC Troubleshooting columns 1989-2024.

Broad peaks medium

Symptom: All peaks on the chromatogram are wider than expected (FWHM > 2× normal)

🔍 Diagnostic tree:
  1. 1. Check whether all peaks are broadened or only some
    → YES: All → instrumental problem (column or system)
    → NO: Only some → chemistry problem (interaction with the column for specific analytes)
  2. 2. Swap in a test column — does the problem disappear?
    → YES: COLUMN worn out — packing damaged, void in the first few mm. Replace it.
    → NO: Problem in the LC system
  3. 3. Check the dead volume — injection loop, connections, detector
    → YES: Loop > 100 µL for a 4.6 mm column or loose connections → replace ferrules, shorten tubing
    → NO: Continue diagnostics
  4. 4. Temperature test: raise the column from 25°C to 40°C
    → YES: Narrower peaks → mass-transfer kinetics too slow (increase T)
    → NO: Continue
  5. 5. Check flow rate vs the optimal van Deemter value for this column
    → YES: Optimum for 4.6mm/5µm = 1.0 mL/min, for 2.1mm/3µm = 0.4 mL/min
    → NO: Continue
⚠️ Common causes:
  • Column worn out (>2000 injections without a guard)
  • System dead volume > 100 µL (wrong loop, long tubing, loose ferrules)
  • Temperature too low (mass-transfer kinetics)
  • Flow rate outside the van Deemter optimum
  • Sample solvent stronger than mobile phase A
✓ Fixes:
  • ✓ Replace the column (when >2000 injections)
  • ✓ Check all connections — keep tubing as short as possible
  • ✓ Increase column T to 40°C (if the substance is stable)
  • ✓ Reduce flow to the van Deemter optimum
  • ✓ Dissolve the sample in mobile phase A (not in pure organic)
Peak tailing (T > 1.5) high

Symptom: Peaks have an extended "tail" on the late-elution side (asymmetry T = b/a > 1.5 per USP)

🔍 Diagnostic tree:
  1. 1. Does the substance contain basic groups (amino, pyridine)?
    → YES: Yes → silanol interactions! Add 0.1% TFA or 5-10 mM TEA to mobile phase A.
    → NO: Continue
  2. 2. Check the mobile-phase pH vs the substance pKa
    → YES: pH = pKa ± 1 → partial ionization, peak split. Move pH ≥ 2 units away from pKa.
    → NO: Continue
  3. 3. Check the column age (>1500 injections?)
    → YES: Yes → exposed silanols (column bleed). Replace with a column with higher endcapping (XTerra, Symmetry).
    → NO: Continue
  4. 4. Does the sample contain metals (Fe, Cu from glass vials)?
    → YES: Yes → use type II clear vials or PFA. Add 0.1mM EDTA to the sample.
    → NO: Continue
⚠️ Common causes:
  • Silanol interactions (basic analyte + silica gel free silanols)
  • pH at the boundary of the analyte pKa (peak split)
  • Old column (column bleed, high silanol activity)
  • Metals in the sample (chelation → tailing)
  • Column overload (>50 µg on a 4.6mm column)
✓ Fixes:
  • ✓ Add 0.1% TFA (UV) or 0.1% formic acid (LC-MS) to mobile phase A
  • ✓ Choose a column with high-purity endcapping: Waters XBridge BEH, Phenomenex Kinetex
  • ✓ Work at pH ≥ 2 units away from pKa
  • ✓ Add 0.1mM EDTA to the sample (Fe/Cu chelation)
  • ✓ Reduce the injection volume to ≤ 20 µL for a 4.6mm column
Baseline drift medium

Symptom: The baseline rises or falls systematically for >5 minutes

🔍 Diagnostic tree:
  1. 1. Are you running a gradient (B% increasing)?
    → YES: Yes → different absorption of phases A vs B at dλ. Solvent change in UV cutoff. Check the % organic UV absorbance.
    → NO: Continue (isocratic)
  2. 2. Check the column temperature — is it stable to ±0.5°C?
    → YES: Yes (stable) → continue
    → NO: Unstable → turn on the column thermostat (>25°C controlled)
  3. 3. Test: turn off the autosampler, run pump+column+detector alone
    → YES: Drift disappears → autosampler contamination (clean the needle, septum)
    → NO: Continue
  4. 4. Check the lamp age (D2 for UV)
    → YES: Yes (>1500 hours) → replace the lamp
    → NO: Continue
⚠️ Common causes:
  • Gradient elution with different UV cutoff of the phases
  • Unstable column T
  • Autosampler contamination of the needle/septum
  • Old UV lamp (>1500h)
  • Detector flow cell fouled
  • Column not equilibrated (<10 column volumes)
✓ Fixes:
  • ✓ Pre-equilibrate the column for 10-15 column volumes at 100% A
  • ✓ Column thermostat on, T 30-40°C stable
  • ✓ Clean the detector flow cell with 50:50 ACN:H2O
  • ✓ Replace the D2 lamp if >1500h
  • ✓ Use baseline subtraction (Chromeleon, Empower native function)
No peak / lost peak critical

Symptom: The expected analyte peak does not appear on the chromatogram

🔍 Diagnostic tree:
  1. 1. Did the injection actually take place?
    → YES: Check the autosampler log, pump pressure (should drop during injection)
    → NO: Autosampler problem → check the loop, needle, sample in the vial
  2. 2. Is the sample in the vial (correct volume, not evaporated)?
    → YES: Continue
    → NO: No sample — re-pipette
  3. 3. Sample stability — prepared >24h ago?
    → YES: Yes → degradation. Re-prepare a fresh sample.
    → NO: Continue
  4. 4. Check the detection wavelength vs the substance λmax
    → YES: Detection at λ does NOT match λmax → no signal. Scan DAD 200-400nm.
    → NO: Continue
  5. 5. Test: inject a pure standard (of known concentration, fresh)
    → YES: The standard gives a peak → problem with the sample (matrix, derivatization)
    → NO: No peak even with the standard → system problem (column, phase, gradient)
⚠️ Common causes:
  • Sample not drawn from the vial (autosampler bug)
  • Sample degraded (>24h pH/temp/light)
  • Detection at the wrong wavelength
  • Wrong mobile phase (e.g. forgotten TFA)
  • Column reversed / wrong stationary phase
  • Substance elutes at the front (V0) → unretained, not visible
✓ Fixes:
  • ✓ Re-prepare a fresh sample per the exact protocol
  • ✓ UV-Vis DAD scan 200-400nm + search for λmax
  • ✓ Check the mobile-phase composition — was TFA added?
  • ✓ Test the reverse column direction (carefully!)
  • ✓ For retention <1 min — lower the % B, MeOH instead of ACN
  • ✓ Check the "expected_rt" table in wp_molgod_hplc_methods
Pressure too high critical

Symptom: Pump pressure > 80% of the column max or system shutdown with a high-pressure error

🔍 Diagnostic tree:
  1. 1. Check that the column is connected correctly (arrow direction)
    → YES: OK
    → NO: Column reversed → flip it (never run it "backwards")
  2. 2. Test: remove the column from the system, run pump+detector alone
    → YES: Pressure drops to <50 bar → problem in the column (clogged)
    → NO: Pressure stays high → in-line filter clogged, frit fouled
  3. 3. Check the pre-column filter (in-line frit)
    → YES: Fouled and brown → replace it
    → NO: Continue
  4. 4. Back-flush the column with 50:50 ACN:H2O without the column — does it disappear?
    → YES: Particles stuck in the first mm — a 30 min flush may recover it
    → NO: Replace the column
⚠️ Common causes:
  • In-line filter (frit) clogged with particles
  • Buffer salting out (precipitation at high %B)
  • Sample contains suspended matter (filter 0.22 µm before injection)
  • Column clogged (column bed compaction)
  • Gradient with a buffer phase + high organic → salt precipitation
✓ Fixes:
  • ✓ ALWAYS filter the sample through 0.22 µm PVDF before injection
  • ✓ Replace the in-line filter every 100 injections (or when pressure rises >20%)
  • ✓ Do NOT use >20mM phosphate buffer + >70% ACN (the salt precipitates)
  • ✓ Flush the column for 30 min with 50:50 ACN:H2O in the reverse direction (when the manufacturer allows it)
  • ✓ Pre-column 4×3mm to protect the main column
Ghost peaks high

Symptom: Unexplained peaks on the chromatogram absent from the calibration

🔍 Diagnostic tree:
  1. 1. Test: blank injection (pure sample solvent)
    → YES: A ghost appears → contamination of the system or eluents
    → NO: Appears only with the sample → matrix
  2. 2. Does the ghost grow with the gradient (elutes at high %B)?
    → YES: Yes → overloaded column or strong-retained from a previous run
    → NO: Independent of the gradient → autosampler carryover
  3. 3. Increase carryover wash (between injections)
    → YES: Helps → carryover was to blame. Use a stronger wash protocol.
    → NO: Continue
  4. 4. Pure water injection — is there a peak?
    → YES: Yes → contamination of the water source (organics from the DI system)
    → NO: Continue
⚠️ Common causes:
  • Carryover in the autosampler needle/loop
  • Eluent contamination (even HPLC-grade)
  • Strong-retained components from previous runs
  • Plastic in the vials (phthalates, PEG from the caps)
  • Insufficiently purified DI water
✓ Fixes:
  • ✓ Strengthen the wash protocol: 100% B → 100% A → 50:50 (3 cycles)
  • ✓ Strong wash: 100% DMSO or 100% MeOH before calibration
  • ✓ Filter the eluents through 0.22 µm PTFE if in doubt
  • ✓ Use amber glass + Teflon-lined caps for samples
  • ✓ Periodic gradient ramp to 100% B for 10 min (clean-out)
📚 Scientific references (Chicago Author-Date) — click to expand
  1. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. Introduction to Modern Liquid Chromatography. 3rd ed. John Wiley & Sons. Chapter 17 (Troubleshooting) pp. 559-616. ISBN 978-0-470-16754-0. https://doi.org/10.1002/9780470508183 [link ↗]
  2. Dolan, John W.. 2014. LC Troubleshooting (monthly column 1989-2024). LCGC North America. [link ↗] — John Dolan 35-letnia seria miesięcznych artykułów problemowych
  3. Kromidas, Stavros. 2017. HPLC Made to Measure: A Practical Handbook for Optimization. 2nd ed. Wiley-VCH. ISBN 978-3-527-31377-1. — Praktyczny przewodnik problem-solving dla labs analitycznych
  4. Dolan, John W.. 2013. When to Modify Method Conditions. 192-199. [link ↗] — Decision flow for changing flow rate / temperature / %B vs swapping columns.
  5. Dong, Michael W.. 2019. HPLC and UHPLC for Practicing Scientists. 2nd ed. Wiley. ISBN 978-1-119-31378-3. https://doi.org/10.1002/9781119313793 [link ↗] — Chapter 9 covers troubleshooting modern UHPLC systems (sub-2 µm particles).
  6. Meyer, Veronika R.. 2010. Practical High-Performance Liquid Chromatography. 5th ed. Wiley. ISBN 978-0-470-68218-0. — Solid step-by-step problem isolation chapter (eluents, columns, instruments).
  7. Snyder, L. R., J. J. Kirkland, and J. L. Glajch. 1997. Practical HPLC Method Development. 2nd ed. Wiley. ISBN 978-0-471-00703-6. — Method-development companion volume with troubleshooting cross-refs.
  8. Carr, Peter W.. 2009. The new physical chemistry of HPLC. 1764-1772. https://doi.org/10.1016/j.chroma.2008.11.094 [link ↗] — Theoretical basis for diagnosing efficiency losses (mass-transfer, eddy diffusion).
  9. Heyden, Yvan Vander, et al.. 2009. Robustness of pharmaceutical liquid chromatographic methods. 2120-2129. https://doi.org/10.1016/j.jchromb.2008.10.052 [link ↗] — How to diagnose method failures vs. system failures (Plackett-Burman).
  10. Engelhardt, Heinz. 2014. 100 Years of Chromatography. 2nd ed. Wiley-VCH. ISBN 978-3-527-33473-5. — Historical context for ghost-peak phenomenology (silica chemistry).
  11. Stoll, Dwight R., and Peter W. Carr. 2017. Two-Dimensional Liquid Chromatography: A State of the Art Tutorial. 519-531. https://doi.org/10.1021/acs.analchem.6b03506 [link ↗]
  12. Carr, Peter W.. 2009. The new physical chemistry of HPLC. 1764-1772. https://doi.org/10.1016/j.chroma.2008.11.094 [link ↗]
  13. Wu, Naijun, and Anton M. Clausen. 2007. Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations. 1167-1182. https://doi.org/10.1002/jssc.200700026 [link ↗]
  14. Engelhardt, Heinz. 2014. 100 Years of Chromatography. 2nd ed. Wiley-VCH. ISBN 978-3-527-33473-5.
  15. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. Probabilistic approach to peak deconvolution in chromatography. 8525-8531. https://doi.org/10.1021/ac101742z [link ↗]
🧪 Solubility and solvent compatibility MolGod_SOLUB_1
Molecule
Boric Acid
Formula
BH3O3
logP
—
Mass (g/mol)
61.84
Polarity
—

⚠️ GC estimate (Hoftyzer–Van Krevelen). No literature HSP data for this CAS — precision ±2 MPa½. Verify experimentally.

Solvent Compat. Ra Visual GC-MS HPLC Applications References
Water (H₂O)− Poor33.3
✗ NieA (aqueous) (RP)
buffercell cultureanalyticalhydrophilic extraction
Ethanol (EtOH)~ Avg.11.6
✗ NieA/B modifier (RP/NP)
extractionspectroscopy (UV-Vis)synthesisHPLC modifier
Methanol (MeOH)− Poor13.7
✗ NieA/B (RP) (RP)
HPLC (eluent)LC-MSKarl FischerUV-transparent do 205 nm
Acetone~ Avg.8.1
✗ NieB modifier (NP)
GC headspacecrystallizationdegreasingsynthesis
Acetonitrile (ACN)− Poor12.3
✗ NieB (RP) (RP)
HPLC eluent (gold standard)LC-MS (wolny cut-off UV 190 nm)peptide analysis
DMSO− Poor14.9
✗ NieN/A (N/A)
NMR (d6-DMSO)cell biology (cryopreservation)drug deliverysynthesis
THF~ Avg.10.4
✗ NieB (NP) (NP)
GPC/SEC (polymer analysis)Grignard synthesisorganometallic
DCM (CH₂Cl₂)− Poor13.4
✓ TakB (NP) (NP)
extractionNP-HPLCGC-MScrystallization (anti-solvent)
Chloroform (CHCl₃)− Poor13.8
✓ TakN/A (toxic) (N/A)
NMR (CDCl3)lipid extraction (Folch method)NP-TLC
Hexane− Poor14.9
✓ TakA (NP) (NP)
NP-HPLCoil extraction (lipids)GC-MSTLC (NP)
Toluene− Poor16.5
✓ TakB (NP) (NP)
NMR (d8-toluene)synthesisDean-Stark azeotropic drying
📚 Scientific references for solvents (Chicago Author-Date) — click to expand

11 solvents × 5 independent scientific sources (NIST/CRC/IARC/Hansen/Reichardt/Smallwood/Wypych/Armarego/Snyder/GESTIS). 55+ full citations below.

Water (H₂O)
  1. NIST — NIST Chemistry WebBook — Water (CAS 7732-18-5)
  2. CRC — CRC Handbook of Chemistry and Physics, 104th ed., Sec. 8 (Properties of Water)
  3. IAPWS — IAPWS Release on Static Dielectric Constant of Water
  4. Reichardt 2011 — Solvents and Solvent Effects in Organic Chemistry
  5. GESTIS — GESTIS Substance Database — Water
Ethanol (EtOH)
  1. NIST — NIST Chemistry WebBook — Ethanol (CAS 64-17-5)
  2. CRC — CRC Handbook — Ethanol physical constants
  3. Snyder & Kirkland — Modern Liquid Chromatography — Ethanol eluotropic
  4. Smallwood — Handbook of Organic Solvent Properties — Ethanol
  5. GESTIS — GESTIS Substance Database — Ethanol
Methanol (MeOH)
  1. NIST — NIST Chemistry WebBook — Methanol (CAS 67-56-1)
  2. CRC — CRC Handbook — Methanol physical constants
  3. Snyder & Kirkland — Modern Liquid Chromatography — MeOH eluotropic, eo=0.95
  4. GESTIS — GESTIS Substance Database — Methanol
Acetone
  1. NIST — NIST Chemistry WebBook — Acetone (CAS 67-64-1)
  2. CRC — CRC Handbook — Acetone physical & thermodynamic constants
  3. Hansen 2007 — Hansen Solubility Parameters — Acetone (dD=15.5, dP=10.4, dH=7.0)
  4. Smallwood — Handbook of Organic Solvent Properties — Acetone
  5. GESTIS — GESTIS Substance Database — Acetone
Acetonitrile (ACN)
  1. NIST — NIST Chemistry WebBook — Acetonitrile (CAS 75-05-8)
  2. CRC — CRC Handbook — Acetonitrile constants
  3. Snyder & Kirkland — Modern Liquid Chromatography — ACN gold-standard HPLC eluent
  4. Reichardt 2011 — Solvents and Solvent Effects — ACN dipolar aprotic
  5. GESTIS — GESTIS Substance Database — Acetonitrile
DMSO
  1. NIST — NIST Chemistry WebBook — DMSO (CAS 67-68-5)
  2. Wypych 2019 — Handbook of Solvents Vol. 1 — DMSO comprehensive properties
  3. Hansen 2007 — HSP — DMSO (dD=18.4, dP=16.4, dH=10.2)
  4. Reichardt 2011 — Solvents and Solvent Effects — DMSO E_T(30)=45.1, dipolar aprotic
  5. GESTIS — GESTIS Substance Database — DMSO
THF
  1. NIST — NIST Chemistry WebBook — THF (CAS 109-99-9)
  2. Armarego 2009 — Purification of Laboratory Chemicals — THF drying & peroxide test
  3. Hansen 2007 — Hansen Solubility Parameters — THF (dD=16.8, dP=5.7, dH=8.0)
  4. Smallwood — Handbook of Organic Solvent Properties — THF
  5. GESTIS — GESTIS Substance Database — Tetrahydrofuran
DCM (CH₂Cl₂)
  1. NIST — NIST Chemistry WebBook — Dichloromethane (CAS 75-09-2)
  2. IARC 71 — IARC Monograph 71 — DCM (Group 2A carcinogen)
  3. Hansen 2007 — Hansen Solubility Parameters — DCM (dD=18.2, dP=6.3, dH=6.1)
  4. Reichardt 2011 — Solvents and Solvent Effects — DCM polarity index
  5. GESTIS — GESTIS Substance Database — Dichloromethane
Chloroform (CHCl₃)
  1. NIST — NIST Chemistry WebBook — Chloroform (CAS 67-66-3)
  2. IARC 73 — IARC Monograph 73 — Chloroform (Group 2B carcinogen)
  3. Hansen 2007 — Hansen Solubility Parameters — CHCl3 (dD=17.8, dP=3.1, dH=5.7)
  4. Reichardt 2011 — Solvents and Solvent Effects — CHCl3 H-bond donor strength
  5. GESTIS — GESTIS Substance Database — Chloroform
n-Hexane
  1. NIST — NIST Chemistry WebBook — n-Hexane (CAS 110-54-3)
  2. ATSDR n-Hexane — ATSDR Toxicological Profile for n-Hexane — neuropatia obwodowa (n-Heksan NIE jest kancerogenem IARC)
  3. Hansen 2007 — Hansen Solubility Parameters — n-Hexane (dD=14.9, dP=0, dH=0)
  4. Snyder & Kirkland — Modern Liquid Chromatography — n-Hexane NP standard, eo=0.00
  5. GESTIS — GESTIS Substance Database — n-Hexane
Toluene
  1. NIST — NIST Chemistry WebBook — Toluene (CAS 108-88-3)
  2. IARC 71 — IARC Monograph 71 — Toluene
  3. Hansen 2007 — Hansen Solubility Parameters — Toluene (dD=18.0, dP=1.4, dH=2.0)
  4. Smallwood — Handbook of Organic Solvent Properties — Toluene
  5. GESTIS — GESTIS Substance Database — Toluene
Solubility theory (applied in compatibility prediction):
  1. Yalkowsky, Samuel H., and Shri C. Valvani. 1980. "Solubility and Partitioning I: Solubility of Nonelectrolytes in Water." Journal of Pharmaceutical Sciences 69 (8): 912–922. https://doi.org/10.1002/jps.2600690814 — General Solubility Equation (GSE): logS = 0.5 − logP − 0.01(MP−25).
  2. Hansen, Charles M. 2007. Hansen Solubility Parameters: A User's Handbook. 2nd ed. CRC Press. https://doi.org/10.1201/9781420006834 — HSP triplet (dD, dP, dH) + Ra formula.
  3. Stefanis, E., and C. Panayiotou. 2008. "Prediction of Hansen Solubility Parameters with a New Group-Contribution Method." Int J Thermophys 29: 568–585. https://doi.org/10.1007/s10765-008-0415-z
  4. Reichardt, Christian, and Thomas Welton. 2011. Solvents and Solvent Effects in Organic Chemistry. 4th ed. Wiley-VCH. https://doi.org/10.1002/9783527632220 — E_T(30) polarity scale, solwatochromia.
  5. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. Introduction to Modern Liquid Chromatography. 3rd ed. Wiley. https://doi.org/10.1002/9780470508183 — Eluotropic series, polarity index.
  6. Van Krevelen, D. W., and K. Te Nijenhuis. 2009. Properties of Polymers. 4th ed. Elsevier. https://doi.org/10.1016/B978-0-08-054819-7.X0001-5 — Hoftyzer–Van Krevelen group contribution dla dD/dP/dH z SMILES.
  7. Marcus, Yizhak. 1998. The Properties of Solvents. Wiley Series in Solution Chemistry, Vol. 4. ISBN 9780471983699 — Complete tabular set of 250+ solvents (ε, μ, donicity, acceptor numbers).
  8. PubChem Compound Database — CAS 10043-35-3 lookup ↗ — logP (XLogP3), water solubility experimental + predicted.

Full bibliography in the REFERENCES accordion (at the bottom of the page) — Chicago Manual of Style 17th ed., Author-Date.

📚 Technical FAQ — Boric Acid (5) MolGod_TECHFAQ_1
❓ Jak przygotować roztwór standardowy kwasu borowego o stężeniu 0.1 M?
Aby przygotować 1 litr roztworu standardowego kwasu borowego (BH3O3) o stężeniu 0.1 M, należy odważyć masę substancji zgodnie z obliczeniem: n = C × V × M, gdzie C = 0.1 mol/L, V = 1 L, M = 61.84 g/mol. Masa kwasu borowego = 0.1 mol/L × 1 L × 61.84 g/mol = 6.184 g. Rozpuścić 6.184 g kwasu borowego w wodzie destylowanej i uzupełnić do objętości 1 litra.
Helpful?
❓ Jak przechowywać kwas borowy, aby zachować jego stabilność?
Kwas borowy należy przechowywać w szczelnie zamkniętych pojemnikach, w temperaturze pokojowej (15-25°C), z dala od światła słonecznego i źródeł ciepła. Wilgotność powietrza nie powinna przekraczać 60%, aby uniknąć absorpcji wilgoci z otoczenia.
Helpful?
❓ Jaka metoda analityczna jest odpowiednia do oznaczania stężenia kwasu borowego w roztworze?
Dla kwasu borowego o masie molowej 61.84 g/mol i logP (wartość nieokreślona, ale zakładając niską lipofilność) zalecana jest metoda HPLC (High-Performance Liquid Chromatography) z detektorem UV. Alternatywnie, można użyć miareczkowania titracją kwasowo-zasadową z fenoloftaleiną jako wskaźnikiem.
Helpful?
❓ Jakie są potencjalne niezgodności chemiczne kwasu borowego?
Kwas borowy może reagować z silnymi utleniaczami (np. nadtlenkiem wodoru) oraz zasadami (tworząc borany). Należy unikać kontaktu z kwasami silniejszymi od niego, które mogą go zneutralizować. Przechowywać oddzielnie od substancji redukujących i mocnych kwasów.
Helpful?
❓ W jakich praktycznych zastosowaniach laboratoryjnych wykorzystuje się kwas borowy?
Kwas borowy jest stosowany jako środek buforujący w zakresie pH 8-9, konserwant w preparatach biologicznych oraz w syntezie organoboranów. W laboratoriach analitycznych używa się go do przygotowania roztworów standardowych i jako odczynnik w miareczkowaniu.
Helpful?
🧮 Laboratory calculators (8) MolGod_LABCALC_1
Dilution (C₁V₁=C₂V₂)
Molarity (M=n/V)
pH Buffer (Henderson-Hasselbalch)
Beer-Lambert (A=εcl)
Mass → Moles
Concentration % → M
ppm → mg/L
Temperature C↔F↔K

Verified formulas: IUPAC Gold Book ↗, DOI ↗

📊 Spectroscopic Databases MolGod_SPECDB_3
📋 Laboratory protocol generator MolGod_PROTOCOL_1

Protocol generated based on: GHS SDS, Aldrich Lab Guide ↗

🏷️ Label generator (QR) MolGod_LABEL_1
Boric acid• BORIC ACID / Orthoboric acid• CAS: 10043-35-3• EC: 233-139-2• Formula: BH3O3• Mass: 61.84 g/molDANGERGHS HAZARD STATEMENTS:H360: May damage fertility or the unborn childP203: Obtain, read and follow all safety instructions before useFOR LABORATORY USE ONLY!DH ScientificScience first. Commerce as consequence.Batch No.: Netto Mass: MFG:
🧪 Solution preparation assistant (Smart Prep) MolGod_PREP_2

Enter what you want to prepare — I'll generate an SOP

Examples below — click to insert:
Preset recipes:
📚 Scientific literature overview — CAS 10043-35-3MolGod_LITHUB_MAIN
⭐ Key findings (scientific literature) 4 publications
🏆 CAS 10043-35-3 — multi-criteria ranking (W12): 30% citations · 20% recency · 20% topic · 15% historical · 15% open access.
  1. #1
    (1987) · National Toxicology Program technical report series
    Why it matters: Selected by multi-criteria score (citations + recency + topic + historical + OA).
    SCORE 3.61 Mechanism Citations: 15 PubMed ↗
  2. #2
    Substance name: Boric acid EC number: 233-139-2 (234-343-4) CAS number: 10043-35-3 (11113-50-1) MEMBER STATE COMMITTEE DRAFT SUPPORT DOCUMENT FOR IDENTIFICATION OF BORIC ACID AS A SUBSTANCE OF VERY HIGH CONCERN BECAUSE OF ITS CMR PROPERTIES
    (2010)
    Why it matters: Selected by multi-criteria score (citations + recency + topic + historical + OA).
    SCORE 2.71 Mechanism Citations: 7
  3. #3
    (2012) · Sax's Dangerous Properties of Industrial Materials
    Why it matters: Selected by multi-criteria score (citations + recency + topic + historical + OA).
    SCORE 0.6 Mechanism DOI ↗
  4. #4
    (2004) · Sax's Dangerous Properties of Industrial Materials
    Why it matters: Selected by multi-criteria score (citations + recency + topic + historical + OA).
    SCORE 0 Mechanism DOI ↗
📈 HPLC gradient — optimizer (LSS) TEMPLATE

logP unknown — PubChem did not return an XLogP value. The gradient below is a generic 5–95% MeCN/H2O template over 15 min; verify parameters before use.

⚠ logP unavailable. PubChem did not return an XLogP3 value for this CAS number. The gradient values below are a generic template — not an LSS fit for this compound.
  • Column: C18
  • Buffer: phosphate
  • Flow: 1 mL/min
  • logP: logP unavailable
  • Ramp: 21% → 95% B, 15 min
  • Total analysis time: 28 min
t (min) %A %B flow (mL/min) Comment
0 79 21 1 start (equilibrium)
2 79 21 1 end of initial hold
17 5 95 1 end of LSS ramp
22 5 95 1 column wash
23 79 21 1 return to init
28 79 21 1 re-equilibration
📚 Scientific references (Chicago Author-Date)
  1. Snyder, Lloyd R., John W. Dolan, and Joseph J. Kirkland. 2010. Introduction to Modern Liquid Chromatography. Wiley. — Chapter 9 — gradient elution, LSS theory (cited as Snyder et al. 2010 in tool description).
  2. Schoenmakers, Peter J. 1986. Optimization of Chromatographic Selectivity: A Guide to Method Development. Elsevier. — Numerical optimization of gradient programs.
  3. Snyder, L. R., and J. W. Dolan. 2007. High-Performance Gradient Elution: The Practical Application of the Linear-Solvent-Strength Model. Wiley. — Foundational LSS reference for the %B_init = 5 + 8·logP heuristic implemented here.
  4. Nikitas, Pavlos, and Adrian Pappa-Louisi. 2009. "Retention models for isocratic and gradient elution in reversed-phase liquid chromatography." Journal of Chromatography A 1216: 1737-1755. [DOI ↗] — Modern review of gradient retention models — basis for non-LSS extensions.
  5. Carr, Peter W.. 2009. "The new physical chemistry of HPLC." Journal of Chromatography A 1216: 1764-1772. [DOI ↗]
  6. Dong, Michael W. 2019. HPLC and UHPLC for Practicing Scientists. Wiley. https://doi.org/10.1002/9781119313793. — Modern UHPLC gradient programming, sub-2 µm scaling rules.
  7. Wu, Naijun, and Anton M. Clausen. 2007. "Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations." Journal of Separation Science 30: 1167-1182. [DOI ↗]
  8. Stoll, Dwight R., and Peter W. Carr. 2017. "Two-Dimensional Liquid Chromatography: A State of the Art Tutorial." Analytical Chemistry 89: 519-531. [DOI ↗] — Reference for orthogonal gradient design (2D-LC second dimension).
  9. Dolan, John W.. 2013. "When to Modify Method Conditions." LCGC North America 31: 192-199.
  10. Meyer, Veronika R. 2010. Practical High-Performance Liquid Chromatography. Wiley. — Chapter 7 — practical gradient design with isokratyczny scouting.
  11. Stoll, Dwight R., and Peter W. Carr. 2017. "Two-Dimensional Liquid Chromatography: A State of the Art Tutorial." Analytical Chemistry 89: 519-531. [DOI ↗]
  12. Carr, Peter W.. 2009. "The new physical chemistry of HPLC." Journal of Chromatography A 1216: 1764-1772. [DOI ↗]
  13. Wu, Naijun, and Anton M. Clausen. 2007. "Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations." Journal of Separation Science 30: 1167-1182. [DOI ↗]
  14. Engelhardt, Heinz. 2014. 100 Years of Chromatography. Wiley-VCH.
  15. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. "Probabilistic approach to peak deconvolution in chromatography." Analytical Chemistry 82: 8525-8531. [DOI ↗]

REST: /wp-json/molgod/v1/hplc/gradient/10043-35-3

📐 HPLC peak symmetry calculator (USP Tf / As)

Calculate the USP tailing factor (Tf) and asymmetry (As) from the peak half-widths. Enter a (left half-width) and b (right half-width) measured at 5% or 10% of peak height.

📚 References (Chicago Author-Date)
  1. USP General Chapter <621>. 2024. "Chromatography." United States Pharmacopeial Convention. [link ↗] — Defines USP Tailing Factor T = (a+b)/(2a) measured at 5% peak height.
  2. International Council for Harmonisation (ICH). 2023. "Validation of Analytical Procedures Q2(R2)." ICH Expert Working Group. [link ↗] — Tailing factor is a system suitability parameter (Section 6).
  3. Foley, Joe P., and John G. Dorsey. 1983. "Equations for calculation of chromatographic figures of merit for ideal and skewed peaks." Analytical Chemistry 55: 730-737 https://doi.org/10.1021/ac00255a033 [link ↗] — Original asymmetry factor As = b/a at 10% height (Foley & Dorsey 1983).
  4. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. "Introduction to Modern Liquid Chromatography." Wiley. https://doi.org/10.1002/9780470508183 [link ↗] — Chapter 2.4 — peak shape diagnostics and remedies.
  5. Dolan, John W.. 2003. "Peak tailing and resolution." LCGC North America 21: 610-614 [link ↗] — How tailing factor degrades effective resolution.
  6. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. "Probabilistic approach to peak deconvolution in chromatography." Analytical Chemistry 82: 8525-8531 https://doi.org/10.1021/ac101742z [link ↗] — Modern numerical deconvolution for asymmetric peaks.
  7. Kromidas, Stavros. 2017. "HPLC Made to Measure: A Practical Handbook for Optimization." Wiley-VCH. — Practical Tf and As thresholds for routine QC.
  8. Dong, Michael W.. 2019. "HPLC and UHPLC for Practicing Scientists." Wiley. https://doi.org/10.1002/9781119313793 [link ↗]
  9. Meyer, Veronika R.. 2010. "Practical High-Performance Liquid Chromatography." Wiley.
  10. Heyden, Yvan Vander, et al.. 2009. "Robustness of pharmaceutical liquid chromatographic methods." Journal of Chromatography B 877: 2120-2129 https://doi.org/10.1016/j.jchromb.2008.10.052 [link ↗]
  11. Stoll, Dwight R., and Peter W. Carr. 2017. "Two-Dimensional Liquid Chromatography: A State of the Art Tutorial." Analytical Chemistry 89: 519-531 https://doi.org/10.1021/acs.analchem.6b03506 [link ↗]
  12. Carr, Peter W.. 2009. "The new physical chemistry of HPLC." Journal of Chromatography A 1216: 1764-1772 https://doi.org/10.1016/j.chroma.2008.11.094 [link ↗]
  13. Wu, Naijun, and Anton M. Clausen. 2007. "Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations." Journal of Separation Science 30: 1167-1182 https://doi.org/10.1002/jssc.200700026 [link ↗]
  14. Engelhardt, Heinz. 2014. "100 Years of Chromatography." Wiley-VCH.
  15. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. "Probabilistic approach to peak deconvolution in chromatography." Analytical Chemistry 82: 8525-8531 https://doi.org/10.1021/ac101742z [link ↗]
📊 Resolution and plate count calculator (Rs, N, H)

Calculate the resolution Rs, the number of theoretical plates N and HETP (H) for a pair of HPLC peaks. Enter the retention times, peak widths (at 50% or at the base) and the column length.

📚 References (Chicago Author-Date)
  1. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. "Introduction to Modern Liquid Chromatography." 3rd ed. John Wiley & Sons. ISBN 978-0-470-16754-0. https://doi.org/10.1002/9780470508183 [link ↗] — Chapter 2 covers resolution, plate count and HETP fundamentals (Snyder et al. 2010).
  2. USP General Chapter <621>. 2024. "Chromatography." USP-NF 2024 ed. United States Pharmacopeial Convention. [link ↗] — Defines Rs >= 1.5 acceptance criterion and N calculation methods.
  3. Dolan, John W.. 2003. "How much resolution is enough?." LCGC North America 21: 350-353 [link ↗] — Practical guidance on Rs targets for routine method development.
  4. Van Deemter, J. J., F. J. Zuiderweg, and A. Klinkenberg. 1956. "Longitudinal diffusion and resistance to mass transfer as causes of nonideality in chromatography." Chemical Engineering Science 5: 271-289 https://doi.org/10.1016/0009-2509(56)80003-1 [link ↗] — Origin of N = 5.54·(tr/w0.5)² half-height plate count formulation.
  5. Giddings, J. Calvin. 1965. "Dynamics of Chromatography, Part I: Principles and Theory." Marcel Dekker. ISBN 978-0-8247-1357-7. — Resolution equation Rs = (1/4)·√N·(α-1)/α·k/(1+k) (master equation).
  6. Foley, Joe P., and John G. Dorsey. 1983. "Equations for calculation of chromatographic figures of merit for ideal and skewed peaks." Analytical Chemistry 55: 730-737 https://doi.org/10.1021/ac00255a033 [link ↗] — Skewed-peak corrections to apparent N.
  7. Knox, John H.. 1977. "Practical aspects of LC theory." Journal of Chromatographic Science 15: 352-364 https://doi.org/10.1093/chromsci/15.9.352 [link ↗]
  8. Carr, Peter W.. 2009. "The new physical chemistry of HPLC." Journal of Chromatography A 1216: 1764-1772 https://doi.org/10.1016/j.chroma.2008.11.094 [link ↗]
  9. Dong, Michael W.. 2019. "HPLC and UHPLC for Practicing Scientists." 2nd ed. Wiley. ISBN 978-1-119-31378-3. https://doi.org/10.1002/9781119313793 [link ↗]
  10. Meyer, Veronika R.. 2010. "Practical High-Performance Liquid Chromatography." 5th ed. Wiley. ISBN 978-0-470-68218-0.
  11. Stoll, Dwight R., and Peter W. Carr. 2017. "Two-Dimensional Liquid Chromatography: A State of the Art Tutorial." Analytical Chemistry 89: 519-531 https://doi.org/10.1021/acs.analchem.6b03506 [link ↗]
  12. Carr, Peter W.. 2009. "The new physical chemistry of HPLC." Journal of Chromatography A 1216: 1764-1772 https://doi.org/10.1016/j.chroma.2008.11.094 [link ↗]
  13. Wu, Naijun, and Anton M. Clausen. 2007. "Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations." Journal of Separation Science 30: 1167-1182 https://doi.org/10.1002/jssc.200700026 [link ↗]
  14. Engelhardt, Heinz. 2014. "100 Years of Chromatography." 2nd ed. Wiley-VCH. ISBN 978-3-527-33473-5.
  15. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. "Probabilistic approach to peak deconvolution in chromatography." Analytical Chemistry 82: 8525-8531 https://doi.org/10.1021/ac101742z [link ↗]
🧪 System Suitability — live calculator (USP <621>)

Enter data from 5–6 injections (areas, tR, tailing, plates) — the calculator computes %RSD and means and checks compliance with USP <621>. You can paste CSV (comma-separated) or edit individual values.

📚 References (Chicago Author-Date)
  1. USP General Chapter <621>. 2024. "Chromatography (System Suitability section)." USP-NF 2024 ed. United States Pharmacopeial Convention. [link ↗] — Defines RSD area < 2%, tailing < 2.0, N > 2000 acceptance criteria.
  2. International Council for Harmonisation (ICH). 2023. "Validation of Analytical Procedures Q2(R2)." ICH Expert Working Group. [link ↗] — Section 5.4 — system suitability is part of method validation.
  3. US Food and Drug Administration (FDA). 2018. "Reviewer Guidance: Validation of Chromatographic Methods." US Food and Drug Administration. [link ↗] — CDER reviewer perspective on chromatographic validation expectations.
  4. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. "Introduction to Modern Liquid Chromatography." 3rd ed. Wiley. — Chapter 2 — system suitability fundamentals (RSD, Tf, N).
  5. Heyden, Yvan Vander, et al.. 2009. "Robustness of pharmaceutical liquid chromatographic methods." — Robustness vs. system suitability — design-of-experiments framework.
  6. Rozet, Eric, et al.. 2013. "Analysis of recent pharmaceutical regulatory documents on analytical method validation."
  7. European Medicines Agency (EMA). 2011. "Guideline on bioanalytical method validation EMEA/CHMP/EWP/192217/2009." EMA. [link ↗] — EMA companion guideline with bioanalytical SS criteria.
  8. Dong, Michael W.. 2019. "HPLC and UHPLC for Practicing Scientists." 2nd ed. Wiley. — UHPLC-specific suitability adjustments (n=5 vs. n=6).
  9. Kazakevich, Yuri V., and Rosario LoBrutto, eds.. 2007. "HPLC for Pharmaceutical Scientists." Wiley-Interscience.
  10. AOAC International. 2016. "Appendix F: Guidelines for Standard Method Performance Requirements." AOAC INTERNATIONAL. [link ↗] — Alternative SS thresholds for food/dietary samples.
  11. Stoll, Dwight R., and Peter W. Carr. 2017. "Two-Dimensional Liquid Chromatography: A State of the Art Tutorial."
  12. Carr, Peter W.. 2009. "The new physical chemistry of HPLC."
  13. Wu, Naijun, and Anton M. Clausen. 2007. "Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations."
  14. Engelhardt, Heinz. 2014. "100 Years of Chromatography." 2nd ed. Wiley-VCH.
  15. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. "Probabilistic approach to peak deconvolution in chromatography."
📋 Status prawny (REACH / TSCA / UK)MolGod_REG_2
JurysdykcjaListaStatusSunset
EUREACH-Ann17listed—
EUSVHClisted—
⚗️ Jonizacja w funkcji pH (Henderson-Hasselbalch)MolGod_PHION_1

Typ: Kwas · pKa: 9.24

024681012140%50%100%% zjonizowany% niejonowypH
pH% jonowy% niejonowy
00.0 %100.0 %
20.0 %100.0 %
40.0 %100.0 %
60.1 %99.9 %
85.4 %94.6 %
1085.2 %14.8 %
1299.8 %0.2 %
14100.0 %0.0 %
Źródła dla tej substancji (6)
  • CRC Handbook 91st ed.
    Lide, David R., ed. 2010. CRC Handbook of Chemistry and Physics. 91st ed. Boca Raton, FL: CRC Press.
  • CRC Handbook 105th ed.
    Haynes, William M., David R. Lide, and Thomas J. Bruno, eds. 2024. CRC Handbook of Chemistry and Physics. 105th ed. Boca Raton, FL: CRC Press.
  • NIST WebBook — link
    Linstrom, Peter J., and William G. Mallard, eds. 2024. NIST Chemistry WebBook. NIST Standard Reference Database Number 69. Gaithersburg, MD: National Institute of Standards and Technology.
  • PubChem CID 7628 — link
    Kim, Sunghwan, Jie Chen, Tiejun Cheng, Asta Gindulyte, Jia He, Siqian He, Qingliang Li, et al. 2023. "PubChem 2023 update." Nucleic Acids Research 51 (D1): D1373-D1380. PubChem CID 7628.
  • IUPAC
    Serjeant, E. P., and Boyd Dempsey. 1979. Ionisation Constants of Organic Acids in Aqueous Solution. IUPAC Chemical Data Series No. 23. Oxford: Pergamon Press.
  • NIST
    Goldberg, Robert N., Nand Kishore, and Rebecca Lennen. 2002. "Thermodynamic Quantities for the Ionization Reactions of Buffers." Journal of Physical and Chemical Reference Data 31 (2): 231-370.
Bibliografia metody (Chicago)
  • Henderson, L. J. 1908. "Concerning the Relationship between the Strength of Acids and Their Capacity to Preserve Neutrality." American Journal of Physiology 21 (4): 173-179.
  • Hasselbalch, K. A. 1917. "Die Berechnung der Wasserstoffzahl des Blutes aus der freien und gebundenen Kohlensäure desselben." Biochemische Zeitschrift 78: 112-144.
  • Po, Henry N., and N. M. Senozan. 2001. "The Henderson-Hasselbalch Equation: Its History and Limitations." Journal of Chemical Education 78 (11): 1499-1503.
  • Avdeef, Alex. 2012. "Absorption and Drug Development: Solubility, Permeability, and Charge State." 2nd ed. Wiley.
  • Avdeef, Alex. 2007. "Solubility of sparingly-soluble ionizable drugs." Advanced Drug Delivery Reviews 59 (7): 568-590.
  • Volgyi, Gergely, et al. 2007. "Potentiometric and spectrophotometric pKa determination of water-insoluble compounds." Analytica Chimica Acta 583 (2): 418-428.
  • Fini, Adamo, Giuseppe Fazio, and Giuseppina Feroci. 1997. "Solubility and solubilization properties of non-steroidal anti-inflammatory drugs." Pharmaceutica Acta Helvetiae 70 (4): 305-318.
  • Mauger, John W., Anthony N. Paruta, and Robert J. Gerraughty. 1972. "Solubilities of sulfadiazine, sulfisomidine, and sulfadimethoxine." Journal of Pharmaceutical Sciences 61 (1): 94-97.
  • Lyman, Warren J., William F. Reehl, and David H. Rosenblatt. 1990. "Handbook of Chemical Property Estimation Methods." American Chemical Society.
  • Marcus, Yizhak. 1998. "The Properties of Solvents." Wiley.
  • Serjeant, E. P., and Boyd Dempsey. 1979. Ionisation Constants of Organic Acids in Aqueous Solution. IUPAC Chemical Data Series No. 23. Oxford: Pergamon Press.
  • Perrin, Douglas D. 1965. Dissociation Constants of Organic Bases in Aqueous Solution. IUPAC. London: Butterworths.
  • Goldberg, Robert N., Nand Kishore, and Rebecca Lennen. 2002. "Thermodynamic Quantities for the Ionization Reactions of Buffers." Journal of Physical and Chemical Reference Data 31 (2): 231-370.
  • Haynes, William M., David R. Lide, and Thomas J. Bruno, eds. 2024. CRC Handbook of Chemistry and Physics. 105th ed. Boca Raton, FL: CRC Press.
  • Lide, David R., ed. 2010. CRC Handbook of Chemistry and Physics. 91st ed. Boca Raton, FL: CRC Press.
  • Kim, Sunghwan, Jie Chen, Tiejun Cheng, Asta Gindulyte, Jia He, Siqian He, Qingliang Li, et al. 2023. "PubChem 2023 update." Nucleic Acids Research 51 (D1): D1373-D1380.
  • Knox, Craig, Mike Wilson, Christen M. Klinger, Mark Franklin, Eponine Oler, Alex Wilson, Allison Pon, et al. 2024. "DrugBank 6.0: the DrugBank Knowledgebase for 2024." Nucleic Acids Research 52 (D1): D1265-D1275.
  • Zdrazil, Barbara, Eloy Felix, Fiona Hunter, Emma J. Manners, James Blackshaw, Sybilla Corbett, Marleen de Veij, et al. 2024. "The ChEMBL Database in 2023." Nucleic Acids Research 52 (D1): D1180-D1192.
  • Kanehisa, Minoru, Miho Furumichi, Yoko Sato, Masayuki Kawashima, and Mari Ishiguro-Watanabe. 2023. "KEGG for taxonomy-based analysis of pathways and genomes." Nucleic Acids Research 51 (D1): D587-D592.
  • Linstrom, Peter J., and William G. Mallard, eds. 2024. NIST Chemistry WebBook. NIST Standard Reference Database Number 69. Gaithersburg, MD: National Institute of Standards and Technology.
  • Nelson, David L., and Michael M. Cox. 2017. Lehninger Principles of Biochemistry. 7th ed. New York: W. H. Freeman.
📤 Embed this molecule on your site

Have a blog, forum or scientific service? Embed the interactive 3D molecule on your website — every reader will see it, with a link below to our shop where they can buy the reagent.

🔗 HTML iframe code (easiest — works everywhere)

Copy and paste into your site's HTML editor:

Adjust width and height to your layout.

⚙ WordPress Shortcode (for other shops with MOL-GOD)

🌐 Direct link (for emails, chats, LinkedIn, Twitter)

LinkedIn Twitter/X Facebook
QR code CAS 10043-35-3

📱 QR code (for printing on flyers / labels / catalogs)

Place it in a product catalog, on a bottle label or a flyer. The customer scans it — sees the 3D molecule on their phone, with a link to your shop.

⬇ Download PNG

🖼 Open Graph image (for social media meta tags)

When you share the link, Facebook/LinkedIn/Discord will automatically fetch the image preview:

Preview
📋 License: The embed keeps a backlink to DH Scientific (required — the shop is the data source). Chemical data comes from PubChem (CC0 — public domain). The embed is FREE for educational, commercial and hobby use.
📚 REFERENCES (Aggregate bibliography, Chicago Author-Date) 107 items

All scientific sources cited in the accordions above for CAS 10043-35-3. Format: Chicago Manual of Style 17th ed., Author-Date system.

🗄️ Scientific databases

  1. PubChem. 2026. PubChem Compound Summary: CAS 10043-35-3. Bethesda, MD: National Center for Biotechnology Information (NCBI), National Library of Medicine. (Accessed 2026-11-12.)
  2. NIST. 2026. NIST Chemistry WebBook: CAS 10043-35-3. Gaithersburg, MD: National Institute of Standards and Technology. https://webbook.nist.gov/cgi/cbook.cgi?ID=10043-35-3. (Accessed 2026-11-12.)
  3. AIST. 2026. Spectral Database for Organic Compounds (SDBS): CAS 10043-35-3. Tsukuba, Japan: National Institute of Advanced Industrial Science and Technology. https://sdbs.db.aist.go.jp/. (Accessed 2026-11-12.)
  4. ECHA. 2026. C&L Inventory and REACH Registration: CAS 10043-35-3. Helsinki: European Chemicals Agency. https://echa.europa.eu/information-on-chemicals/annex-vi-to-clp. (Accessed 2026-11-12.)

📐 Standards / Guidelines

  1. ICH. 2026. "ICH Harmonised Guideline: CAS 10043-35-3." Geneva: International Council for Harmonisation. https://database.ich.org/sites/default/files/Q1A%28R2%29%20Guideline.pdf. (Accessed 2026-11-12.)
  2. National Fire Protection Association (NFPA). 2024. "NFPA 30: Flammable and Combustible Liquids Code." NFPA, Quincy, MA. https://www.nfpa.org/codes-and-standards/all-codes-and-standards/list-of-codes-and-standards/detail?code=30.
  3. Occupational Safety and Health Administration (OSHA). 2023. "29 CFR 1910.106 — Flammable Liquids." U.S. Department of Labor, Federal Register. https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.106.
  4. European Chemicals Agency (ECHA). 2024. "Annex VI to Regulation (EC) No 1272/2008 (CLP) — Harmonised Classification and Labelling." ECHA, Helsinki / Official Journal of the European Union. https://echa.europa.eu/regulations/clp/clp-classification.
  5. European Committee for Standardization (CEN). 2016. "EN 374-1:2016 — Protective gloves against dangerous chemicals and micro-organisms — Part 1: Terminology and performance requirements for chemical risks." CEN, Brussels. https://standards.cencenelec.eu/dyn/www/f?p=205:110:::::FSP_PROJECT,FSP_ORG_ID:38536,6080&cs=1B0DAA8B85DF42E4A2C70E5D71F0BFA32.
  6. European Committee for Standardization (CEN). 2001. "EN 166:2001 — Personal eye-protection — Specifications." CEN, Brussels. https://standards.cencenelec.eu/dyn/www/f?p=CEN:110:0::::FSP_PROJECT:6541&cs=1F1A4E0A78C4DB6A28DBE2E8C29D89DCF.
  7. European Committee for Standardization (CEN). 2009. "EN 14605:2005+A1:2009 — Protective clothing against liquid chemicals — Performance requirements for clothing with liquid-tight (Type 3) or spray-tight (Type 4) connections." CEN, Brussels. https://standards.cencenelec.eu/dyn/www/f?p=CEN:110:0::::FSP_PROJECT:21581&cs=1A04A2D3C7CC58E9E6CB58D55F7EBFB7E.
  8. National Institute for Occupational Safety and Health (NIOSH). 2017. "Recommendations for Chemical Protective Clothing: A Companion to the NIOSH Pocket Guide." U.S. Department of Health & Human Services / CDC. https://www.cdc.gov/niosh/ncpc/default.html.
  9. Occupational Safety and Health Administration (OSHA). 2011. "Personal Protective Equipment — General requirements." U.S. Department of Labor — 29 CFR 1910.132. https://www.osha.gov/laws-regs/regulations/standardnumber/1910/1910.132.

📖 Books

  1. Hansen, Charles M. 2007. Hansen Solubility Parameters: A User's Handbook, 2nd ed.. Boca Raton, FL: CRC Press. https://www.routledge.com/Hansen-Solubility-Parameters-A-Users-Handbook/Hansen/p/book/9780849372483.
  2. Barton, Allan F. M. 1991. CRC Handbook of Solubility Parameters and Other Cohesion Parameters: 2nd ed.. Boca Raton, FL: CRC Press. https://www.routledge.com/CRC-Handbook-of-Solubility-Parameters-and-Other-Cohesion-Parameters/Barton/p/book/9780849301766.
  3. Connors, Kenneth A., Gordon L. Amidon, and Valentino J. Stella. 1986. Chemical Stability of Pharmaceuticals: A Handbook for Pharmacists, 2nd ed.. New York: Wiley. https://doi.org/10.1002/0471734683.
  4. Rumble, John R., ed. 2019. CRC Handbook of Chemistry and Physics: 100th Edition. Boca Raton, FL: CRC Press. https://hbcp.chemnetbase.com/.
  5. Urben, Peter G. 2017. Bretherick's Handbook of Reactive Chemical Hazards, 8th Edition. Academic Press / Elsevier, Oxford. https://www.sciencedirect.com/book/9780081010594.

📘 Monographs

  1. IARC. 2026. IARC Monographs on the Identification of Carcinogenic Hazards to Humans: CAS 10043-35-3. Lyon, France: International Agency for Research on Cancer, World Health Organization. (Accessed 2026-11-12.)

📄 Scientific articles (peer-reviewed)

  1. Stefanis, Emmanuel, and Costas Panayiotou. 2008. "Prediction of Hansen Solubility Parameters with a New Group-Contribution Method." International Journal of Thermophysics 29: 568-585. https://doi.org/10.1007/s10765-008-0415-z.
  2. Stoll, Vincent S., and John S. Blanchard. 1990. "Buffers: Principles and Practice: In Methods in Enzymology, vol. 182." San Diego: Academic Press. https://doi.org/10.1016/0076-6879(90)82008-P.

🌐 Websites

  1. ECHA. 2023. "Guidance on the Application of the CLP Criteria." European Chemicals Agency. https://echa.europa.eu/guidance-documents/guidance-on-clp.
  2. European Parliament. 2006. "Regulation (EC) No 1907/2006 (REACH)." Official Journal of the European Union L 396: 1–849.
  3. ECHA. 2023. "Candidate List of Substances of Very High Concern for Authorisation." European Chemicals Agency. https://echa.europa.eu/candidate-list-table.
  4. European Parliament. 2008. "Regulation (EC) No 1272/2008 on Classification, Labelling and Packaging of Substances and Mixtures (CLP)." Official Journal of the European Union L 353: 1–1355.
  5. ECHA. 2017. "Guidance on the Compilation of Safety Data Sheets." Version 3.1. European Chemicals Agency. ECHA-17-G-01-EN. https://echa.europa.eu/documents/10162/23047722/sds_en.pdf.
  6. ECHA. 2022. "Restrictions Under REACH — Annex XVII." European Chemicals Agency. https://echa.europa.eu/substances-restricted-under-reach.
  7. United Nations. 2021. Globally Harmonized System of Classification and Labelling of Chemicals (GHS). 9th revised ed. ST/SG/AC.10/30/Rev.9. New York and Geneva: United Nations. https://unece.org/ghs-rev9-2021.
  8. ECHA. 2020. "Understanding REACH." European Chemicals Agency. https://echa.europa.eu/regulations/reach/understanding-reach.
  9. U.S. OSHA. 2026. "Occupational Safety and Health Guidance: CAS 10043-35-3." Washington, DC: U.S. Occupational Safety and Health Administration. https://www.osha.gov/hazwoper. (Accessed 2026-11-12.)
  10. National Fire Protection Association (2018) — NFPA 472: Standard for Competence of Responders to Hazardous Materials/Weapons of Mass Destruction Incidents https://www.nfpa.org/codes-and-standards/nfpa-472.
  11. European Union. 2026. "EUR-Lex Legal Database: CAS 10043-35-3." Luxembourg: Publications Office of the European Union. https://eur-lex.europa.eu/legal-content/EN/TXT/?uri=celex:32012L0018. (Accessed 2026-11-12.)
  12. U.S. National Institute for Occupational Safety and Health (2024) — NIOSH Pocket Guide to Chemical Hazards https://www.cdc.gov/niosh/npg/.
  13. Snyder, Lloyd R., John W. Dolan, and Joseph J. Kirkland. 2010. Introduction to Modern Liquid Chromatography. Wiley.
  14. Schoenmakers, Peter J.. 1986. Optimization of Chromatographic Selectivity: A Guide to Method Development. Elsevier.
  15. Snyder, L. R., and J. W. Dolan. 2007. High-Performance Gradient Elution: The Practical Application of the Linear-Solvent-Strength Model. Wiley.
  16. Nikitas, Pavlos, and Adrian Pappa-Louisi. 2009. "Retention models for isocratic and gradient elution in reversed-phase liquid chromatography." Journal of Chromatography A 1216: 1737-1755. https://doi.org/10.1016/j.chroma.2008.10.005.
  17. Carr, Peter W.. 2009. "The new physical chemistry of HPLC." Journal of Chromatography A 1216: 1764-1772. https://doi.org/10.1016/j.chroma.2008.11.094.
  18. Dong, Michael W.. 2019. HPLC and UHPLC for Practicing Scientists. Wiley. https://doi.org/10.1002/9781119313793.
  19. Wu, Naijun, and Anton M. Clausen. 2007. "Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations." Journal of Separation Science 30: 1167-1182. https://doi.org/10.1002/jssc.200700026.
  20. Dolan, John W.. 2013. "When to Modify Method Conditions." LCGC North America 31: 192-199. https://www.chromatographyonline.com/view/when-modify-method-conditions.
  21. Meyer, Veronika R.. 2010. Practical High-Performance Liquid Chromatography. Wiley.
  22. Engelhardt, Heinz. 2014. 100 Years of Chromatography. Wiley-VCH.
  23. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. "Probabilistic approach to peak deconvolution in chromatography." Analytical Chemistry 82: 8525-8531. https://doi.org/10.1021/ac101742z.
  24. Van Deemter, J. J., F. J. Zuiderweg, and A. Klinkenberg. 1956. "Longitudinal diffusion and resistance to mass transfer as causes of nonideality in chromatography." https://doi.org/10.1016/0009-2509(56)80003-1.
  25. Giddings, J. Calvin. 1965. "Dynamics of Chromatography, Part I: Principles and Theory." Marcel Dekker.
  26. Poppe, Hans. 1997. "Some reflections on speed and efficiency of modern chromatographic methods." https://doi.org/10.1016/S0021-9673(97)00376-2.
  27. Wu, Naijun, and Anton M. Clausen. 2007. "Fundamental and practical aspects of ultrahigh pressure liquid chromatography for fast separations." https://doi.org/10.1002/jssc.200700026.
  28. Carr, Peter W.. 2009. "The new physical chemistry of HPLC." https://doi.org/10.1016/j.chroma.2008.11.094.
  29. Knox, John H.. 1977. "Practical aspects of LC theory." https://doi.org/10.1093/chromsci/15.9.352.
  30. Snyder, L. R., J. J. Kirkland, and J. L. Glajch. 1997. "Practical HPLC Method Development." Wiley.
  31. Vivó-Truyols, Gabriel, and Hans-Gerd Janssen. 2010. "Probabilistic approach to peak deconvolution in chromatography." https://doi.org/10.1021/ac101742z.
  32. Sadek, Paul C.. 2002. "The HPLC Solvent Guide." Wiley-Interscience.
  33. Snyder, L. R.. 1978. "Classification of the solvent properties of common liquids." https://doi.org/10.1093/chromsci/16.6.223.
  34. Reichardt, Christian, and Thomas Welton. 2010. "Solvents and Solvent Effects in Organic Chemistry." Wiley-VCH.
  35. Vailaya, Anant, and Csaba Horváth. 1998. "Retention thermodynamics in hydrophobic interaction chromatography." https://doi.org/10.1021/ie980212h.
  36. Krstulović, Andrea M., and Phyllis R. Brown. 1981. "Reversed-phase High-Performance Liquid Chromatography." Wiley.
  37. Boysen, Reinhard I., and Milton T. W. Hearn. 2009. "Multi-modal HPLC of proteins." https://doi.org/10.1093/chromsci/47.8.645.
  38. USP General Chapter <621>. 2024. "Chromatography." United States Pharmacopeial Convention. https://www.usp.org/harmonization-standards/pdg/general-chapters/chromatography.
  39. Foley, Joe P., and John G. Dorsey. 1983. "Equations for calculation of chromatographic figures of merit for ideal and skewed peaks." https://doi.org/10.1021/ac00255a033.
  40. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. "Introduction to Modern Liquid Chromatography." Wiley. https://doi.org/10.1002/9780470508183.
  41. Dolan, John W.. 2003. "Peak tailing and resolution." https://www.chromatographyonline.com/view/peak-tailing-and-resolution.
  42. Kromidas, Stavros. 2017. "HPLC Made to Measure: A Practical Handbook for Optimization." Wiley-VCH.
  43. Heyden, Yvan Vander, et al.. 2009. "Robustness of pharmaceutical liquid chromatographic methods." https://doi.org/10.1016/j.jchromb.2008.10.052.
  44. United States Pharmacopeial Convention. 2024. "USP <621> Chromatography." In United States Pharmacopeia and National Formulary, USP 47-NF 42. Rockville, MD: USP. https://www.uspnf.com/.
  45. European Pharmacopoeia Commission. 2024. "2.2.46 Chromatographic Separation Techniques." In European Pharmacopoeia, 11th ed. Strasbourg: Council of Europe — EDQM. https://www.edqm.eu/en/european-pharmacopoeia-ph-eur-11th-edition.
  46. International Organization for Standardization. 2017. "ISO/IEC 17025:2017 General Requirements for the Competence of Testing and Calibration Laboratories." Geneva: ISO. https://www.iso.org/standard/66912.html.
  47. Kolthoff, Izaak Maurits, and Philip J. Elving, eds. 1978. Treatise on Analytical Chemistry, Part I: Theory and Practice. 2nd ed. New York: Wiley-Interscience.
  48. Skoog, Douglas A., F. James Holler, and Stanley R. Crouch. 2018. Principles of Instrumental Analysis. 7th ed. Boston: Cengage Learning.
  49. Christian, Gary D., Purnendu K. Dasgupta, and Kevin A. Schug. 2014. Analytical Chemistry. 7th ed. Hoboken, NJ: Wiley.
  50. EURACHEM/CITAC. 2012. "Quantifying Uncertainty in Analytical Measurement." 3rd ed. EURACHEM/CITAC Guide CG 4. https://www.eurachem.org/images/stories/Guides/pdf/QUAM2012_P1.pdf.
  51. Snyder, Lloyd R., Joseph J. Kirkland, and John W. Dolan. 2010. "Introduction to Modern Liquid Chromatography." John Wiley & Sons. https://doi.org/10.1002/9780470508183.
  52. Dolan, John W.. 2003. "How much resolution is enough?." https://www.chromatographyonline.com/view/how-much-resolution-enough.
  53. USP General Chapter <621>. 2024. "Chromatography (System Suitability section)." United States Pharmacopeial Convention. https://www.usp.org/harmonization-standards/pdg/general-chapters/chromatography.
  54. U.S. FDA. 2026. "FDA Guidance and Resources: CAS 10043-35-3." Silver Spring, MD: U.S. Food and Drug Administration. https://www.fda.gov/media/74954/download. (Accessed 2026-11-12.)
  55. Rozet, Eric, et al.. 2013. "Analysis of recent pharmaceutical regulatory documents on analytical method validation." https://doi.org/10.1016/j.chroma.2007.03.111.
  56. European Medicines Agency (EMA). 2011. "Guideline on bioanalytical method validation EMEA/CHMP/EWP/192217/2009." EMA. https://www.ema.europa.eu/en/documents/scientific-guideline/guideline-bioanalytical-method-validation_en.pdf.
  57. Kazakevich, Yuri V., and Rosario LoBrutto, eds.. 2007. "HPLC for Pharmaceutical Scientists." Wiley-Interscience. https://doi.org/10.1002/9780470087954.
  58. AOAC International. 2016. "Appendix F: Guidelines for Standard Method Performance Requirements." AOAC INTERNATIONAL. https://www.aoac.org/wp-content/uploads/2019/08/app_f.pdf.
  59. International Organization for Standardization. 1994. "ISO 5725-2:1994 Accuracy (Trueness and Precision) of Measurement Methods and Results — Part 2: Basic Method for the Determination of Repeatability and Reproducibility of a Standard Measurement Method." Geneva: ISO. https://www.iso.org/standard/11834.html.
  60. Grubbs, Frank E. 1950. "Sample Criteria for Testing Outlying Observations." Annals of Mathematical Statistics 21 (1): 27–58.
  61. Dixon, Wilfrid J. 1950. "Analysis of Extreme Values." Annals of Mathematical Statistics 21 (4): 488–506.
  62. Snedecor, George W., and William G. Cochran. 1989. Statistical Methods. 8th ed. Ames, IA: Iowa State University Press.
  63. Student [William Sealy Gosset]. 1908. "The Probable Error of a Mean." Biometrika 6 (1): 1–25.
  64. International Organization for Standardization. 2005. "ISO 3534-1:2006 Statistics — Vocabulary and Symbols — Part 1: General Statistical Terms and Terms Used in Probability." Geneva: ISO. https://www.iso.org/standard/40145.html.
  65. Thompson, Michael, Stephen L. R. Ellison, and Roger Wood. 2002. "Harmonized Guidelines for Single-Laboratory Validation of Methods of Analysis." Pure and Applied Chemistry 74 (5): 835–855.
  66. United Nations Economic Commission for Europe. 2024. European Agreement Concerning the International Carriage of Dangerous Goods by Road (ADR), Applicable as from 1 January 2025 (ECE/TRANS/352). Geneva: UNECE. https://unece.org/transport/dangerous-goods/adr-2025-edition.
  67. International Air Transport Association. 2026. Dangerous Goods Regulations (DGR). 67th ed. Montreal: IATA. https://www.iata.org/en/programs/cargo/dgr/.
  68. International Maritime Organization. 2024. International Maritime Dangerous Goods (IMDG) Code, 2024 Edition (Amendment 42-24). London: IMO. https://www.imo.org/en/OurWork/Safety/Pages/DangerousGoods-default.aspx.
  69. United Nations. 2025. Recommendations on the Transport of Dangerous Goods: Model Regulations (Orange Book). 24th revised ed. ST/SG/AC.10/1/Rev.24. New York and Geneva: United Nations. https://unece.org/transport/dangerous-goods/un-model-regulations.
  70. International Civil Aviation Organization. 2025. Technical Instructions for the Safe Transport of Dangerous Goods by Air (Doc 9284). 2025–2026 ed. Montreal: ICAO. https://www.icao.int/safety/DangerousGoods/Pages/technical-instructions.aspx.
  71. United States Pharmacopeia (USP) Convention. 2024. "USP General Chapter <621> Chromatography." USP. https://www.usp.org/sites/default/files/usp/document/harmonization/gen-chapter/g05_pf_30_4_2004.pdf.
  72. European Medicines Agency (EMA). 2011. "Guideline on bioanalytical method validation EMEA/CHMP/EWP/192217/2009." EMA Committee for Medicinal Products for Human Use. https://www.ema.europa.eu/en/documents/scientific-guideline/guideline-bioanalytical-method-validation_en.pdf.
  73. Sejm RP. 2026. "Internetowy System Aktow Prawnych (ISAP): CAS 10043-35-3." Warszawa: Kancelaria Sejmu Rzeczypospolitej Polskiej. https://isap.sejm.gov.pl/isap.nsf/DocDetails.xsp?id=WDU20200000010. (Accessed 2026-11-12.)
  74. Ministerstwo Klimatu i Srodowiska. 2026. "Baza Danych o Odpadach (BDO): CAS 10043-35-3." Warszawa: Ministerstwo Klimatu i Srodowiska. https://bdo.mos.gov.pl/. (Accessed 2026-11-12.)
  75. Pohanish, Richard P.. 2017. "Sittig's Handbook of Toxic and Hazardous Chemicals and Carcinogens." Elsevier.
  76. Lewis, Richard J.. 2012. "Sax's Dangerous Properties of Industrial Materials." Wiley.
  77. NIOSH. 2024. "Pocket Guide to Chemical Hazards." U.S. Department of Health and Human Services. https://www.cdc.gov/niosh/npg/.
  78. United Nations Economic Commission for Europe (UNECE). 2023. "European Agreement concerning the International Carriage of Dangerous Goods by Road (ADR 2025)." UNECE. https://unece.org/transport/standards/transport/dangerous-goods/adr-2025.
  79. IPCS INCHEM. 2024. "International Programme on Chemical Safety — Waste Management Guidelines." WHO/UNEP/ILO. https://www.inchem.org/.
  80. International Organization for Standardization. 2017. "ISO/IEC 17025:2017 General requirements for the competence of testing and calibration laboratories." ISO. https://www.iso.org/standard/66912.html.
  81. World Health Organization. 2010. "WHO Good Manufacturing Practices for Pharmaceutical Products: Main Principles (WHO Technical Report Series No. 957, Annex 3)." WHO Press. https://www.who.int/publications/m/item/trs957-annex3.
  82. European Medicines Agency. 2014. "Guideline on Process Validation for Finished Products — Information and Data to Be Provided EMA/CHMP/CVMP/QWP/BWP/70278/2012." European Medicines Agency. https://www.ema.europa.eu/en/documents/scientific-guideline/guideline-process-validation-finished-products-information-data-be-provided-regulatory-submissions-revision-1_en.pdf.
  83. United States Pharmacopeial Convention. 2024. "United States Pharmacopeia and National Formulary, USP 47-NF 42." USP. https://www.uspnf.com/.
  84. European Pharmacopoeia Commission. 2024. "European Pharmacopoeia 11th Edition." Council of Europe — EDQM. https://www.edqm.eu/en/european-pharmacopoeia-ph-eur-11th-edition.
  85. Pharmaceutical Inspection Co-operation Scheme (PIC/S). 2021. "Guide to Good Manufacturing Practice for Medicinal Products PE 009-15." PIC/S Secretariat, Geneva. https://picscheme.org/en/publications.
  86. International Pharmaceutical Excipients Council (IPEC) and Pharmaceutical Quality Group (PQG). 2017. "Joint IPEC-PQG Good Manufacturing Practices Guide for Pharmaceutical Excipients." IPEC-Americas. https://ipecamericas.org/sites/default/files/IPECPQGGMPGuide2017.pdf.
📥 Download BibTeX📥 Download RISImport into Zotero/Mendeley/EndNote/Papers.